Department of Physical Therapy, New York Institute of Technology, Old Westbury, NY, USA.
J Man Manip Ther. 2021 Jun;29(3):176-180. doi: 10.1080/10669817.2020.1805691. Epub 2020 Aug 18.
The Cervical Range of Motion (CROM) device is a valid and reliable clinical tool used to measure full cervical rotation, however, its reliability for measuring upper cervical rotation is unknown. Assess between-week test-retest reliability of the CROM device in measuring upper cervical rotation Thirty students participated in this test-retest reliability study. The CROM device was used to measure left and right cervical rotation in both a seated neutral and fully flexed head-neck position. Interclass correlation coefficient (ICC) was calculated for all motions. Measurement error was determined using standard error of measurement (SEM) and minimal detectable change (MDC). The CROM device demonstrated moderate to good reliability (ICCs 0.65-0.9) of full and upper cervical rotation. The SEMs and MDCs of this study are small and suggest that the chance of repeated measurement error was relatively minimal for the between-week trials. The CROM device is a reliable outcome tool for measuring upper cervical rotation. The clinical implications of these findings suggest that therapists can utilize the CROM device to more completely examine all planes of upper and full cervical mobility. It may also assist in identifying upper cervical ROM limitations associated with underlying cervical pathology or motion dysfunction.
颈椎活动度(CROM)设备是一种有效且可靠的临床工具,用于测量整个颈椎旋转,但它测量上颈椎旋转的可靠性尚不清楚。评估 CROM 设备在上颈椎旋转测量中的两周间测试-再测试可靠性 30 名学生参加了这项测试-再测试可靠性研究。CROM 设备用于测量坐姿中立位和完全头颈部屈曲位时的左右颈椎旋转。计算了所有运动的组内相关系数(ICC)。使用测量误差(SEM)和最小可检测变化(MDC)确定测量误差。CROM 设备在上颈椎和全颈椎旋转的测量中具有中等至良好的可靠性(ICC 0.65-0.9)。本研究的 SEM 和 MDC 较小,表明在两周间的试验中,重复测量误差的可能性相对较小。CROM 设备是测量上颈椎旋转的可靠结果工具。这些发现的临床意义表明,治疗师可以使用 CROM 设备更全面地检查上颈椎和全颈椎活动的所有平面。它还可能有助于确定与潜在颈椎病理学或运动功能障碍相关的上颈椎 ROM 限制。