The use of electromagnetic tracking technology for measurement of passive cervical range of motion: a pilot study.

作者信息

Morphett Adrian Lindsay, Crawford Colin M, Lee Don

机构信息

School of Science and Engineering, Swinbourne University of Technology, Melbourne, Australia.

出版信息

J Manipulative Physiol Ther. 2003 Mar-Apr;26(3):152-9. doi: 10.1016/S0161-4754(02)54107-5.

Abstract

OBJECTIVE

To investigate the use of electromagnetic tracking technology for measurement of passive cervical range of motion (ROM).

DESIGN

Passive cervical ranges of motion from one extreme to the other were measured in 3 planes (transverse, frontal, and sagittal) by using an electromagnetic tracking system (ETS), the cervical range of motion device (CROM), and visual estimation (VE) with 2 blinded examiners, using a repeated measures design.

SETTING

Swinburne University of Technology, Australia.

INTERVENTION

Four studies were undertaken: (1) Measurement of ROM using 3 methods: the CROM, the ETS, and VE. Two examiners conducted measurements. (2) Measurement of ROM with the CROM and the ETS simultaneously by 1 examiner. (3) Measurement of ROM with the ETS by 2 examiners. (4) Measurement of medium-term reliability of ROM using the ETS over a 24-hour period by 1 examiner.

MAIN OUTCOME MEASURE

The intraclass correlation coefficient (ICC) was used to investigate the reliability within and between each method and reliability of ROM over a 24-hour period.

RESULTS

Study 1: Intra-instrument ICCs ranged from fair to high for the 3 measurement methods. The ETS performed best. Inter-instrument ICCs were poor. Study 2: Direct comparison of the ETS and the CROM yielded high ICCs for rotation and flexion/extension and fair for lateral flexion. Study 3: Interexaminer ICCs using the ETS were high for rotation, good for lateral flexion, and fair for flexion/extension; intraexaminer reliability was high for all planes. Study 4: 24-hour reliability of ROM using the ETS was good for rotation and lateral flexion and poor for flexion/extension.

CONCLUSION

The ETS used in this investigation was an accurate instrument and efficient method for measurement and recording of passive cervical ROM. The ETS had high intraexaminer and fair-to-high interexaminer reliability for the measurement of extremes of ROM in 3 planes. It is probable that differences in the experience levels of examiners did affect interexaminer reliability of the ETS. The ETS and the CROM compared well in rotation and flexion/extension.

摘要

目的

探讨电磁跟踪技术在测量颈椎被动活动范围(ROM)中的应用。

设计

采用重复测量设计,由两名不知情的检查者使用电磁跟踪系统(ETS)、颈椎活动范围测量装置(CROM)和视觉估计法(VE),在三个平面(横向、额状面和矢状面)测量颈椎从一个极端位置到另一个极端位置的被动活动范围。

地点

澳大利亚斯威本科技大学。

干预措施

进行了四项研究:(1)使用三种方法测量ROM:CROM、ETS和VE。两名检查者进行测量。(2)由一名检查者同时使用CROM和ETS测量ROM。(3)由两名检查者使用ETS测量ROM。(4)由一名检查者在24小时内使用ETS测量ROM的中期可靠性。

主要观察指标

组内相关系数(ICC)用于研究每种方法内部和之间的可靠性以及24小时内ROM的可靠性。

结果

研究1:三种测量方法的仪器内ICC范围从中等到高。ETS表现最佳。仪器间ICC较差。研究2:ETS和CROM的直接比较显示,旋转和屈伸的ICC较高,侧屈的ICC中等。研究3:使用ETS的检查者间ICC在旋转时较高,侧屈时良好,屈伸时中等;检查者内可靠性在所有平面均较高。研究4:使用ETS测量的ROM在24小时内的可靠性在旋转和侧屈时良好,屈伸时较差。

结论

本研究中使用的ETS是一种准确的仪器和有效的方法,用于测量和记录颈椎被动ROM。ETS在测量三个平面ROM极端值时具有较高的检查者内可靠性和中等至高的检查者间可靠性。检查者经验水平的差异可能确实影响了ETS的检查者间可靠性。ETS和CROM在旋转和屈伸方面比较相似。