Suppr超能文献

一种用于同时定量唾液褪黑素和皮质醇的灵敏且特异的液相色谱-串联质谱分析方法:开发和与免疫测定法的比较。

A Sensitive and Specific Liquid Chromatography-Tandem Mass Spectrometry Assay for Simultaneous Quantification of Salivary Melatonin and Cortisol: Development and Comparison With Immunoassays.

机构信息

Department of Laboratory Medicine and Genetics, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, Korea.

Department of Neurology, Ilsan Paik Hospital, Inje University College of Medicine, Goyang, Korea.

出版信息

Ann Lab Med. 2021 Jan;41(1):108-113. doi: 10.3343/alm.2021.41.1.108. Epub 2020 Aug 25.

Abstract

Melatonin and cortisol are clinically important for diagnosing sleep and mood disorders. We developed and validated a liquid chromatography-tandem mass spectrometry (LC-MS/MS) assay for simultaneous measurement of salivary melatonin and cortisol concentrations according to the Clinical and Laboratory Standards Institute guidelines. Additionally, we compared the LC-MS/MS assay with immunoassays, ELISA (Direct Salivary Melatonin Elisa EK-DSM, Bühlmann Laboratories AG, Schönenbuch, Switzerland) and electrochemiluminescence immunoassay (Cortisol II, Roche, Mannheim, Germany), using 121 saliva samples. The LC-MS/MS assay exhibited good performance in terms of linearity, precision, accuracy, limit of detection, lower limit of quantification, extraction recovery, carry-over, and matrix effect. The LC-MS/MS assay and immunoassays showed strong correlation (Pearson's r=0.910 for melatonin, r=0.955 for cortisol), but demonstrated a significant mean bias of 23.2% (range 54.0-143.7%) for melatonin and 48.9% (range 59.7-184.7%) for cortisol. Our LC-MS/MS assay provided more sensitive and reliable salivary melatonin and cortisol quantification results compared with immunoassays.

摘要

褪黑素和皮质醇在诊断睡眠和情绪障碍方面具有重要的临床意义。我们根据临床和实验室标准协会的指南,开发并验证了一种同时测量唾液褪黑素和皮质醇浓度的液相色谱-串联质谱(LC-MS/MS)检测方法。此外,我们使用 121 份唾液样本,将 LC-MS/MS 检测方法与免疫测定法(ELISA,Direct Salivary Melatonin Elisa EK-DSM,Bühlmann Laboratories AG,Schönenbuch,Switzerland)和电化学发光免疫测定法(Cortisol II,Roche,Mannheim,Germany)进行了比较。LC-MS/MS 检测方法在线性、精密度、准确度、检测限、定量下限、提取回收率、交叉污染和基质效应方面表现出良好的性能。LC-MS/MS 检测方法和免疫测定法之间具有很强的相关性(褪黑素的 Pearson r=0.910,皮质醇的 r=0.955),但褪黑素的平均偏差为 23.2%(范围 54.0-143.7%),皮质醇的平均偏差为 48.9%(范围 59.7-184.7%)。与免疫测定法相比,我们的 LC-MS/MS 检测方法提供了更敏感和可靠的唾液褪黑素和皮质醇定量结果。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/cc6e/7443521/4570df922c15/alm-2021-41-1-108f1.jpg

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验