J Diabetes Sci Technol. 2021 Sep;15(5):1076-1083. doi: 10.1177/1932296820949930. Epub 2020 Aug 17.
Globally millions of people with diabetes still prick their fingers to measure blood glucose. The aim of this study was to comprehensively evaluate and to compare three lancing devices set at the minimum ("1") and at the maximum ("5") lancing depth with respect to blood volume (BV) and pain related to lancing.
Lancing devices tested were A-, B- (both: HTL-Strefa S.A., Poland), and C- (Ascensia Diabetes Care, Switzerland), all used with personal lancets of three sizes 28G, 30G, and 33G. BVs were measured with calibrated capillaries. Pain related to lancing was expressed as a derivative of pain rating with visual analog scale.
In 90 participants with diabetes, 360 lancing procedures were performed. Overall, BV and pain were higher for "maximum" compared to "minimum" lancing depth (for both < .001). Pain differed between devices ( ≤ .001), overall was higher for device A compared to B or C; in paired comparisons differences were significant for the following settings: A > B for 28G/1 and 33G/1, B > C for 30G/1, and A > C for 28G/1, 30G/1, and 33G/1. In aggregated comparison we did not prove a significant effect of lancet size on either BV nor pain ( = .1109, = .4966, respectively).
BV depended mainly on lancing depth. Pain depended on lancing depth and to some degree on device type. The results may serve as a source of comparative data of lancing devices performance for studies in which other lancing devices and/or lancets would be tested.The study was registered at ClinicalTrials.gov: NCT03479619.
全球有数百万糖尿病患者仍需刺破手指来测量血糖。本研究旨在全面评估和比较三种刺血仪在最低(“1”)和最高(“5”)穿刺深度时的采血体积(BV)和穿刺相关疼痛。
测试的刺血仪分别为 A-、B-(均为:HTL-Strefa S.A.,波兰)和 C-(Ascensia Diabetes Care,瑞士),均与三种规格的个人采血针(28G、30G 和 33G)一起使用。BV 用校准的毛细管测量。穿刺相关疼痛用视觉模拟评分的导数表示。
在 90 名糖尿病患者中,共进行了 360 次穿刺操作。总体而言,与最低穿刺深度相比,最高穿刺深度时 BV 和疼痛更高(均<0.001)。不同设备之间的疼痛存在差异(≤0.001),总体而言,A 设备比 B 或 C 设备更高;在配对比较中,以下设置存在显著差异:A 比 B 在 28G/1 和 33G/1 时更高,B 比 C 在 30G/1 时更高,而 A 比 C 在 28G/1、30G/1 和 33G/1 时更高。在综合比较中,我们没有证明针的大小对 BV 或疼痛有显著影响(分别为=0.1109,=0.4966)。
BV 主要取决于穿刺深度。疼痛取决于穿刺深度,在一定程度上取决于设备类型。这些结果可以作为其他刺血仪和/或采血针测试的比较数据来源,用于研究中。本研究在 ClinicalTrials.gov 注册:NCT03479619。