硬脑膜替代物在台式模型中对超声评估超声透光性颅成形术的成像质量产生不同的干扰。

Dural Substitutes Differentially Interfere with Imaging Quality of Sonolucent Transcranioplasty Ultrasound Assessment in Benchtop Model.

机构信息

Department of Neurosurgery, Baylor College of Medicine, Houston, Texas, USA.

Department of Neurosurgery, Baylor College of Medicine, Houston, Texas, USA.

出版信息

World Neurosurg. 2020 Dec;144:e389-e394. doi: 10.1016/j.wneu.2020.08.166. Epub 2020 Sep 2.

Abstract

BACKGROUND

Sonolucent cranioplasty implants were recently introduced into clinical practice and tested for use with transcranioplasty ultrasound (TCUS). In situations where dural substitutes (DSs) are needed during closure, such as in extracranial-intracranial bypass, it is unclear if the DS influences the imaging quality of TCUS. The aim of this study was to assess the influence of DSs on imaging quality during TCUS assessment with sonolucent cranioplasty.

METHODS

A tofu vascular brain model was constructed with a ClearFit implant in between 2 layers of tofu. Injection of saline solution through a 6-F Envoy DA XB endovascular catheter (6-F Envoy DA XB) inserted into the deeper tofu layer mimicked an intracranial vessel. TCUS image quality, including Doppler, with 4 different DSs (DuraGen, Durepair, DuraMatrix-Onlay, DuraMatrix Suturable) placed under the cranioplasty was compared against a control by 2 examiners. A literature search of MEDLINE and EMBASE was conducted to find previous reports of acoustic properties of DSs.

RESULTS

TCUS assessment including Doppler was feasible with the model in longitudinal and axial planes, and the pulsatile particle injection was visualized in real time. DuraGen and Durepair showed Doppler quality and picture detail comparable to the control, while the DuraMatrix-Onlay and DuraMatrix Suturable were inferior to the control. The literature search yielded only 1 previous report on acoustic properties of DSs.

CONCLUSIONS

DSs interfere differentially with imaging quality during TCUS assessment. However, these results are based on a benchtop model and need to be further assessed in the clinical setting.

摘要

背景

超声可透过型颅骨修补植入物最近被引入临床实践,并经过颅超声(TCUS)测试。在需要硬脑膜替代品(DS)进行闭合的情况下,例如在颅外-颅内旁路手术中,尚不清楚 DS 是否会影响 TCUS 的成像质量。本研究旨在评估 DS 对超声可透过型颅骨修补物 TCUS 评估中成像质量的影响。

方法

在 2 层豆腐之间构建了一个带有 ClearFit 植入物的豆腐血管大脑模型。通过插入深层豆腐层的 6-F Envoy DA XB 血管内导管(6-F Envoy DA XB)注入生理盐水,模拟颅内血管。通过 2 名检查者比较 4 种不同的 DS(DuraGen、Durepair、DuraMatrix-Onlay、DuraMatrix Suturable)放置在颅骨修补物下时与对照相比的 TCUS 图像质量,包括多普勒。通过 MEDLINE 和 EMBASE 进行文献检索,以查找关于 DS 声学特性的先前报告。

结果

该模型的纵向和轴向平面均可行 TCUS 评估,包括多普勒,并实时可视化脉动粒子注射。DuraGen 和 Durepair 的多普勒质量和图像细节与对照相当,而 DuraMatrix-Onlay 和 DuraMatrix Suturable 则不如对照。文献检索仅得出 1 篇关于 DS 声学特性的先前报告。

结论

DS 在 TCUS 评估期间对成像质量产生不同程度的干扰。然而,这些结果基于台式模型,需要在临床环境中进一步评估。