Monofrequency electrical impedance mammography (EIM) diagnostic system in breast cancer screening.

机构信息

Unidad de Investigación en Epidemiología Clínica, Unidad Médica de Alta Especialidad No. 1 Bajío, Instituto Mexicano del Seguro Social, B. López Mateos Esq. Insurgentes S/N, Colonia: Los Paraísos, CP 37320, Leon, Guanajuato, Mexico.

Servicio de Radiología e Imagen Unidad Médica de Alta Especialidad No. 48, Instituto Mexicano del Seguro Social, Leon, Guanajuato, Mexico.

出版信息

BMC Cancer. 2020 Sep 14;20(1):876. doi: 10.1186/s12885-020-07283-5.

Abstract

BACKGROUND

Some evidence has shown that malignant breast tumours have lower electrical impedance than surrounding normal tissues. Electrical impedance could be used as an indicator for breast cancer detection. The purpose of our study was to analyse the sensitivity and specificity of electrical impedance mammography (EIM) and its implementation for the differential diagnosis of pathological lesions of the breast, either alone or in combination with mammography/ultrasound, in 1200 women between 25 and 70 years old.

METHODS

This study is a prospective, cross-sectional epidemiological observational study of serial screening. The women were invited to participate and signed a consent letter. Impedance imaging of the mammary gland was evaluated with the computerized mammography equipment of MEIK electroimpedance v.5.6. (0.5 mA, 50 kHz), developed and manufactured by PKF SIM-Technika®. The successful identification of breast cancer along with the sensitivity, specificity, and positive and negative predictive values of EIM were determined as follows: % sensitivity; % specificity; % positive predictive value (PPV); and % negative predictive value (NPV).

RESULTS

EIM had a sensitivity of 85% and a specificity of 96%; the positive predictive value was 12%, and the negative predictive value was 99%. Seven cases were biopsy confirmed cancers. Significant correlations between the electrical conductivity index and body mass index (BMI) (p = 0.04) and patient age were observed (p = 0.01). We also observed that the average conductivity distribution increased according to age group (p = 0.001). We used the chi-squared test to assess the interactions between percent density and BMI (normal < 25 kg/m2 (n = 310), overweight 25-29.9 kg/m2 (n = 418) and obese ≥30 (n = 437)) (p <  0.05). The patients with a diagnosis of mammary carcinoma had a BMI of 35.51 kg/m2.

CONCLUSIONS

Our results demonstrate that the use of monofrequency electrical impedance mammography (EIM) in the detection of breast cancer had a sensitivity and specificity of 85 and 96%, respectively. These findings may support future research in the early detection of breast cancer. EIM is a non-radiation method that may also be used as a screening method for young women with dense breasts and a high risk of developing breast cancer.

摘要

背景

一些证据表明,恶性乳腺肿瘤的电阻抗值低于周围正常组织。电阻抗可作为乳腺癌检测的指标。本研究的目的是分析电阻抗乳腺成像(EIM)的敏感性和特异性,以及其单独或与乳腺 X 线摄影/超声联合用于诊断乳腺病理病变的能力,共纳入 1200 名 25 至 70 岁的女性。

方法

这是一项前瞻性、横断面的系列筛查的观察性研究。邀请女性参加并签署知情同意书。使用 MEIK 电阻抗 v.5.6 计算机乳腺成像设备(0.5 mA,50 kHz)进行乳腺阻抗成像,由 PKF SIM-Technika®研发和制造。通过 EIM 成功识别乳腺癌,以及 EIM 的敏感性、特异性、阳性预测值和阴性预测值如下:%敏感性;%特异性;%阳性预测值(PPV);%阴性预测值(NPV)。

结果

EIM 的敏感性为 85%,特异性为 96%;阳性预测值为 12%,阴性预测值为 99%。7 例经活检证实为癌症。观察到电阻抗指数与体重指数(BMI)(p=0.04)和患者年龄之间存在显著相关性(p=0.01)。我们还观察到,根据年龄组,平均电导率分布增加(p=0.001)。我们使用卡方检验评估了密度百分比与 BMI(正常<25 kg/m2(n=310),超重 25-29.9 kg/m2(n=418)和肥胖≥30 kg/m2(n=437)(p<0.05)之间的相互作用。诊断为乳腺癌的患者的 BMI 为 35.51 kg/m2。

结论

我们的结果表明,使用单频电阻抗乳腺成像(EIM)检测乳腺癌的敏感性和特异性分别为 85%和 96%。这些发现可能支持乳腺癌早期检测的进一步研究。EIM 是一种非放射性方法,也可用于检测乳房致密且乳腺癌风险较高的年轻女性。