Desai Dharmin D, Johnson E L, Cordrey Ivan L
Department of Radiation Oncology, CHI Memorial Hospital, Chattanooga, TN, USA.
Department of Radiation Medicine, University of Kentucky Chandler Medical Center, Lexington, KY, USA.
J Appl Clin Med Phys. 2020 Nov;21(11):278-282. doi: 10.1002/acm2.13026. Epub 2020 Sep 30.
In stereotactic body radiation therapy (SBRT), R50% is a common metric for intermediate dose spill and is defined in RTOG 0915 as the ratio of 50% isodose cloud volume (IDC50%) to the planning target volume (PTV). By coupling sound physical principles with the basic definition of intermediate dose spill, we derive an exact analytical expression for R50% for the case of a spherical volume. This expression for R50% depends on three quantities: the surface area of PTV (SA ), the volume of PTV (V ), and the dose gradient Δr. Validity of our analytical expression for R50% was confirmed via direct comparison to peer-reviewed, multi-institutional, diverse clinical data. The comparison of our R50% values computed from our analytical expression to the clinical data yielded an average percent difference of 3.8 ± 4.5%.
在立体定向体部放射治疗(SBRT)中,R50%是用于衡量中等剂量泄漏的常用指标,在RTOG 0915中被定义为50%等剂量云体积(IDC50%)与计划靶体积(PTV)的比值。通过将合理的物理原理与中等剂量泄漏的基本定义相结合,我们推导出了球形体积情况下R50%的精确解析表达式。这个R50%的表达式取决于三个量:PTV的表面积(SA)、PTV的体积(V)和剂量梯度Δr。通过与经过同行评审的多机构多样临床数据直接比较,证实了我们R50%解析表达式的有效性。将根据我们的解析表达式计算出的R50%值与临床数据进行比较,得出的平均百分比差异为3.8±4.5%。