Tear Film Osmolarity Measurement in Japanese Dry Eye Patients Using a Handheld Osmolarity System.

作者信息

Shimazaki Jun, Sakata Miki, Den Seika, Iwasaki Miki, Toda Ikuko

机构信息

Department of Ophthalmology, Tokyo Dental College, Ichikawa General Hospital, Chiba 272-8513, Japan.

Shinjuku Eye Clinic, Tokyo 169-0074, Japan.

出版信息

Diagnostics (Basel). 2020 Oct 5;10(10):789. doi: 10.3390/diagnostics10100789.

Abstract

We studied the efficacy and safety of a handheld osmolarity measurement system (I-PEN) in Japanese patients with dry eye disease (DED) and non-DED subjects. In this prospective, multicenter study, tear osmolarity was examined using the I-PEN in a total of 122 eyes divided into DED ( = 71) and non-DED ( = 51) groups. Subjective symptoms were assessed using the Dry Eye-Related Quality-of-Life Score (DEQS) questionnaire. Ocular surface condition was evaluated in terms of fluorescein tear breakup time (FBUT) and tear breakup pattern (TBUP), and by fluorescein staining and Schirmer's test. The I-PEN measurements were performed safely in the majority of cases. There was no statistically significant difference in mean tear film osmolarity between the DED and non-DED groups (294.76 ± 16.39 vs. 297.76 ± 16.72 mOsms/L, respectively, = 0.32). No significant correlations were observed between osmolarity values and DEQS score, FBUT, or the Schirmer score. Osmolarity did not differ among TBUP subgroups. This prospective clinical study found no correlations between the tear film osmolarity values obtained with the I-PEN system and any subjective or objective parameters of DED. Further studies are required to determine the utility of the I-PEN system in other settings.

摘要

我们研究了一种手持式渗透压测量系统(I-PEN)在日本干眼症(DED)患者和非DED受试者中的疗效和安全性。在这项前瞻性多中心研究中,使用I-PEN对总共122只眼睛进行了泪液渗透压检查,这些眼睛被分为DED组(n = 71)和非DED组(n = 51)。使用干眼相关生活质量评分(DEQS)问卷评估主观症状。通过荧光素泪膜破裂时间(FBUT)、泪膜破裂模式(TBUP)、荧光素染色和施密特试验评估眼表状况。在大多数情况下,I-PEN测量操作安全。DED组和非DED组之间的平均泪膜渗透压无统计学显著差异(分别为294.76±16.39和297.76±16.72 mOsms/L,P = 0.32)。渗透压值与DEQS评分、FBUT或施密特评分之间未观察到显著相关性。不同TBUP亚组之间的渗透压无差异。这项前瞻性临床研究发现,I-PEN系统获得的泪膜渗透压值与DED的任何主观或客观参数之间均无相关性。需要进一步研究以确定I-PEN系统在其他情况下的实用性。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/a264/7601915/006990e957e0/diagnostics-10-00789-g001.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验