Department of Otolaryngology-HNS and Communicative Disorders, University of Louisville, Louisville, Kentucky, USA.
Department of Communication Sciences and Disorders, Syracuse University, Syracuse, New York, USA.
Ear Hear. 2021 May/Jun;42(3):531-546. doi: 10.1097/AUD.0000000000000954.
Because unresolved debris in the ear canal or middle ear of newborns may produce high false positive rates on hearing screening tests, it has been suggested that an outer/middle ear measure can be included at the time of hearing screening. A potential measure is power absorbance (absorbance), which indicates the proportion of power in a broadband acoustic stimulus that is absorbed through the outer/middle ear. Although absorbance is sensitive to outer/middle dysfunction at birth, there is large variability that limits its accuracy. Acoustic leaks caused by poor probe fitting further exacerbate this issue. The objectives of this work were to: (1) develop criteria to indicate whether a change in absorbance occurs in association with probe fit; (2) describe the variability in absorbance due to poor fitting; and (3) evaluate test-retest variability with probe reinsertions, excluding poor fits.
An observational cross-sectional design was used to evaluate changes in absorbance due to probe fit and probe reinsertion. Repeated measurements were recorded in 50 newborns (98 ears) who passed TEOAE screenings and were <48 hours of age. One absorbance measurement was chosen as the baseline that served as a best-fit reference in each ear. Changes in absorbance, called absorbance probe-fit Δ, were calculated relative to the baseline in each ear. Correlations were assessed between the absorbance probe-fit Δ and low-frequency absorbance, impedance magnitude, impedance phase, and equivalent volume, to determine which measures predicted poor fits. Criteria were derived from the strongest of these correlations and their performance was analyzed. Next, measurements with poor/leaky fits were identified, and the changes in absorbance that they introduced were analyzed. Excluding the poor fits, test-retest differences in absorbance, called reinsertion Δ, were determined. Variability was assessed using the SDs associated with absorbance, absorbance probe-fit Δ, and reinsertion Δ.
Based on the analysis of 12 moderate-strong correlations, the following criteria were adopted to identify measurements with poor fits: (1) impedance phase-based criterion (500 to 1000 Hz) > -0.11 cycles and (2) absorbance-based criterion (250 to 1000 Hz) > 0.58. Poor-fit measurements introduced statistically significant increases in absorbance up to 0.1 for 1000 to 6000 Hz, and up to 0.4 for frequencies <1000 Hz. Reinsertion Δ were ≤0.02, and were significant for 500 to 5000 Hz. The SDs of absorbance probe-fit Δ were greatest and similar to overall absorbance SD in the low frequencies. Separately, the SDs of reinsertion Δ were also greatest and similar to low-frequency absorbance SD.
Poor probe fits introduced the greatest inflation in absorbance for frequencies < 500 Hz, and a smaller but significant inflation for higher frequencies, consistent with controlled experiments on acoustic leaks in adults. Importantly, inflation of absorbance in diagnostically sensitive 1000 to 2000 Hz may impact its clinical performance. Test-retest with probe reinsertion contributed significantly to absorbance variability, especially in the low frequencies, consistent with reports in adults, even though changes were smaller than those associated with poor probe fit. The results indicate that variability in absorbance was reduced by minimizing acoustic leaks. Pending further validation, the probe-fit criteria developed in this work can be recommended to ensure proper probe fit.
由于新生儿耳道或中耳中未解决的碎片可能导致听力筛查测试中出现高假阳性率,因此有人建议在外耳/中耳检查时可以加入一项中耳/外耳测量。一种潜在的测量方法是功率吸收率(吸收),它表示宽带声刺激中被外耳/中耳吸收的功率比例。虽然吸收率在出生时对外耳/中耳功能障碍很敏感,但由于存在很大的变异性,其准确性受到限制。探头贴合不良引起的声漏进一步加剧了这个问题。本研究的目的是:(1)制定标准,以指示吸收率是否因探头贴合而发生变化;(2)描述因探头贴合不良而导致的吸收率的变异性;(3)评估探头重新插入时的测试-再测试变异性,排除不良贴合。
本研究采用观察性横断面设计来评估探头贴合和探头重新插入对吸收率的影响。在通过 TEOAE 筛查且年龄小于 48 小时的 50 名新生儿(98 只耳朵)中进行了重复测量。在每只耳朵中选择一个吸收率测量值作为基线,作为最佳拟合参考。吸收率的变化,称为吸收率探头贴合 Δ,相对于每个耳朵的基线进行计算。评估吸收率探头贴合 Δ与低频吸收率、阻抗幅度、阻抗相位和等效体积之间的相关性,以确定哪些指标可以预测探头贴合不良。从这些相关性中得出最强的标准,并分析其性能。接下来,确定存在不良贴合/泄漏的探头,分析它们引入的吸收率变化。排除不良贴合的探头,确定吸收率的再插入 Δ。使用与吸收率、吸收率探头贴合 Δ和再插入 Δ相关的标准差来评估变异性。
基于对 12 个中等强度相关性的分析,采用以下标准来识别探头贴合不良的测量值:(1)基于阻抗相位的标准(500 至 1000 Hz)>-0.11 个周期,和(2)基于吸收率的标准(250 至 1000 Hz)>0.58。不良贴合探头引入的吸收率变化在 1000 至 6000 Hz 之间最大可达 0.1,在<1000 Hz 之间最大可达 0.4。再插入 Δ最大可达 0.02,在 500 至 5000 Hz 之间具有统计学意义。吸收率探头贴合 Δ的标准差最大,在低频段与总体吸收率的标准差相似。此外,再插入 Δ的标准差也最大,与低频吸收率的标准差相似。
不良探头贴合会导致<500 Hz 频率的吸收率显著增加,而在更高频率下则会导致较小但显著的增加,这与成人的声漏控制实验结果一致。重要的是,1000 至 2000 Hz 这一诊断敏感频率的吸收率增加可能会影响其临床性能。探头重新插入的测试-再测试对吸收率的变异性有显著贡献,尤其是在低频段,这与成人的报告一致,尽管变化比探头贴合不良引起的变化要小。结果表明,通过最大限度地减少声漏,可以降低吸收率的变异性。在进一步验证之前,本研究中制定的探头贴合标准可以被推荐以确保探头贴合良好。