肾皮质血流的电影CT测量

Cine-CT measurement of cortical renal blood flow.

作者信息

Jaschke W R, Gould R G, Cogan M G, Sievers R, Lipton M J

机构信息

Department of Radiology, University of California at San Francisco 94143.

出版信息

J Comput Assist Tomogr. 1987 Sep-Oct;11(5):779-84. doi: 10.1097/00004728-198709000-00007.

Abstract

A modified indicator-dilution technique using radiographic contrast material and a cine-CT scanner was used to measure blood flow in the renal cortex of dogs. To validate this technique, CT measurements were correlated with simultaneous measurements of flow determined by radioactive microspheres. Measurements were taken during euvolemic conditions and after hemorrhage. Thirty-nine measurements were compared, covering a flow range from 1 to 7 ml min-1 g-1, and a good correlation was found between the cine-CT and microsphere results (r = 0.93; p less than 0.001). Additionally, cine-CT measurements were made of the mean transit time (MTT) of contrast material through the renal cortex, and the reciprocal of these MTT values was also well correlated to microsphere determined flow (r = 0.94; p less than 0.001). Thus, cine-CT appears to be a promising new technique for measuring renal blood flow.

摘要

采用一种改良的指示剂稀释技术,利用放射造影剂和电影CT扫描仪来测量犬肾皮质的血流量。为验证该技术,将CT测量结果与通过放射性微球同时测定的血流量进行相关性分析。测量在血容量正常状态下及出血后进行。共比较了39次测量结果,血流量范围为1至7毫升/分钟·克-1,电影CT与微球测量结果之间存在良好的相关性(r = 0.93;p < 0.001)。此外,还通过电影CT测量了造影剂通过肾皮质的平均通过时间(MTT),这些MTT值的倒数也与微球测定的血流量高度相关(r = 0.94;p < 0.001)。因此,电影CT似乎是一种很有前景的测量肾血流量的新技术。