De Bie J, Diemberger I, Mason J W
Mortara Instrument Europe s.r.l., Bologna, Italy.
Department of Experimental, Diagnostic and Specialty Medicine, University of Bologna, Bologna, Italy.
J Electrocardiol. 2020 Nov-Dec;63:75-82. doi: 10.1016/j.jelectrocard.2020.10.006. Epub 2020 Oct 21.
Electrocardiograph-generated measurements of PR, QRS, and QT intervals are generally thought to be more precise than manual measurements on paper records. However, the performance of different programs has not been well compared.
Routinely obtained digital electrocardiograms (ECGs), including over 500 pediatric ECGs, were used to create over 2000 10 s analog ECGs that were replayed through seven commercially available electrocardiographs. The measurements for PR interval, QRS duration, and QT interval made by each program were extracted and compared against each other (using the median of the programs after correction for program bias) and the population mean values.
Small but significant systematic biases were seen between programs. The smallest and largest variation from the population mean differed by 4.7 ms for PR intervals, 5.8 ms for QRS duration, and 12.4 ms for QT intervals. In pairwise comparison programs showed similar accuracy for most ECGs, with the average absolute errors at the 75th percentile for PR intervals being 4-6 ms from the median, QRS duration 4-8 ms, and QT interval 6-10 ms. However, substantial differences were present in the numbers and extent of large, clinically significant errors (e.g at the 98th percentile), for which programs differed by a factor of two for absolute errors, as well as differences in the mix of overestimations and underestimations.
When reading digital ECGs, users should be aware that small systematic differences exist between programs and that there may be large clinically important errors in difficult cases.
通常认为心电图仪生成的PR、QRS和QT间期测量值比纸质记录上的手动测量更精确。然而,不同程序的性能尚未得到很好的比较。
使用常规获取的数字心电图(ECG),包括500多份儿科心电图,创建了2000多个10秒的模拟心电图,通过七台商用心电图仪进行回放。提取每个程序对PR间期、QRS时限和QT间期的测量值,并相互比较(使用校正程序偏差后各程序的中位数)以及总体平均值。
各程序之间存在微小但显著的系统偏差。与总体平均值的最小和最大差异,PR间期为4.7毫秒,QRS时限为5.8毫秒,QT间期为12.4毫秒。在成对比较中,大多数心电图的程序显示出相似的准确性,PR间期第75百分位数处的平均绝对误差与中位数相差4 - 6毫秒,QRS时限为4 - 8毫秒,QT间期为6 - 10毫秒。然而,在数量和程度上存在大量具有临床意义的大误差(例如在第98百分位数处),各程序的绝对误差相差两倍,并且高估和低估的组合也存在差异。
在解读数字心电图时,用户应意识到不同程序之间存在微小的系统差异,并且在困难病例中可能存在具有重要临床意义的大误差。