Suppr超能文献

提高重症护理点血糖仪结果质量的有效干预措施。

Effective interventions to improve the quality of critically high point-of-care glucose meter results.

作者信息

Shaw Julie Lv, McCudden Christopher R, Colantonio David A, Booth Ronald A, Lin Danny C, Blasutig Ivan M, Moran Thomas, Trofimczuk Dana, Carriere Chantal, Gharra Anas, Portelance Cheryl, Tremblay Chantal, Dupaul Dianne, Breton Nathalie, Angelkovski Merina, Jariwala Chetan, Embleton Maureen, Campbell Christine, Groulx Kristina, Larmour Karen

机构信息

Eastern Ontario Regional Laboratories Association, Canada.

Division of Biochemistry, The Ottawa Hospital, Canada.

出版信息

Pract Lab Med. 2020 Oct 19;22:e00184. doi: 10.1016/j.plabm.2020.e00184. eCollection 2020 Nov.

Abstract

OBJECTIVES

Point-of-care testing (POCT) is testing performed outside the traditional laboratory, often at the patient bedside. In hospital settings, blood glucose is the most common POCT. Staff performing POCT are not usually laboratory trained; they are clinical staff with a primary focus on treating patients. Clinical staff find POCT quality assurance (QA) practices burdensome and are often non-compliant. In hospitals within EORLA (Eastern Ontario Regional Laboratories Association), all critically high POCT glucose results must be repeated prior to acting, according to policy. Compliance with this policy is audited regularly.

DESIGN

and methods: All POCT glucose tests performed in participating sites between January and June 2018 and June and December 2019 were audited for compliance with the critical repeat policy. The discordant repeat rate was also determined for each audit period. Between January and May 2019, there were interventions aimed at improving compliance with the repeat policy.

RESULTS

Compliance with the critical repeat policy increased from 30 to 57% in 2019 compared to 2018, following nursing education and implementation of notifications on the glucose meters themselves. The rate of discordant repeat results (>20% different from initial) also improved at most sites in 2019 compared to 2018. Nurses cited insufficient cleaning of patient hands prior to initial testing as the primary reason for discordant repeats.

CONCLUSIONS

Operator compliance with POCT QA policies is an ongoing challenge requiring continual audit, feedback and education. A strong POCT multi-disciplinary committee with supports from senior and clinical leadership in an organization are key to improving compliance.

摘要

目的

即时检验(POCT)是在传统实验室之外进行的检测,通常在患者床边进行。在医院环境中,血糖检测是最常见的即时检验项目。进行即时检验的工作人员通常未接受过实验室培训;他们是主要专注于治疗患者的临床工作人员。临床工作人员发现即时检验质量保证(QA)措施繁琐,且常常不遵守规定。在东安大略地区实验室协会(EORLA)所属的医院中,根据政策,所有即时检验得出的极高血糖结果在采取行动之前都必须重复检测。会定期对该政策的遵守情况进行审核。

设计与方法

对2018年1月至6月以及2019年6月至12月期间参与研究的各站点所进行的所有即时检验血糖检测进行审核,以检查是否符合关键重复检测政策。还确定了每个审核期间的不一致重复率。在2019年1月至5月期间,采取了旨在提高对重复检测政策遵守情况的干预措施。

结果

与2018年相比,2019年在进行了护理教育并在血糖仪上实施通知后,对关键重复检测政策的遵守率从30%提高到了57%。与2018年相比,2019年大多数站点不一致重复结果(与初始结果相差>20%)的发生率也有所改善。护士们指出,初次检测前对患者手部清洁不足是不一致重复检测的主要原因。

结论

操作人员对即时检验质量保证政策的遵守是一项持续存在的挑战,需要持续进行审核、反馈和教育。一个强大的即时检验多学科委员会,并得到组织内高层和临床领导的支持,是提高遵守率的关键。