Int J Lab Hematol. 2021 Jun;43(3):372-377. doi: 10.1111/ijlh.13380. Epub 2020 Nov 4.
The Kleihauer-Betke (KB) test is the diagnostic standard for the quantification of fetomaternal hemorrhage (FMH). Manual analysis of KB slides suffers from inter-observer and inter-laboratory variability and low efficiency. Flow cytometry provides accurate quantification of FMH with high efficiency but is not available in all hospitals or at all times. We have developed an automated KB counting system that uses machine learning to identify and distinguish fetal and maternal red blood cells (RBCs). In this study, we aimed to evaluate and compare the accuracy, precision, and efficiency of the automated KB counting system with manual KB counting and flow cytometry.
The ratio of fetal RBCs of the same blood sample was quantified by manual KB counting, automated KB counting, and flow cytometry, respectively. Forty patients were enrolled in this comparison study.
Comparing the automated KB counting system with flow cytometry, the mean bias in measuring the ratio of fetal RBCs was 0.0048%, with limits of agreement ranging from -0.22% to 0.23%. Using flow cytometry results as a benchmark, results of automated KB counting were more accurate than those from manual counting, with a lower mean bias and narrower limits of agreement. The precision of automated KB counting was higher than that of manual KB counting (intraclass correlation coefficient 0.996 vs 0.79). The efficiency of automated KB counting was 200 times that of manual counting by the certified technologists.
Automated KB counting provides accurate and precise FMH quantification results with high efficiency.
Kleihauer-Betke(KB)检测是定量胎儿母体出血(FMH)的诊断标准。KB 玻片的手动分析存在观察者间和实验室间的可变性和低效率的问题。流式细胞术具有高效、准确的 FMH 定量能力,但并非所有医院或所有时间都能提供。我们开发了一种自动化 KB 计数系统,该系统使用机器学习来识别和区分胎儿和母体的红细胞(RBC)。在这项研究中,我们旨在评估和比较自动化 KB 计数系统与手动 KB 计数和流式细胞术的准确性、精密度和效率。
分别通过手动 KB 计数、自动化 KB 计数和流式细胞术来定量同一血样中胎儿 RBC 的比例。这项比较研究共纳入了 40 名患者。
将自动化 KB 计数系统与流式细胞术进行比较,测量胎儿 RBC 比例的平均偏差为 0.0048%,一致性界限范围为-0.22%至 0.23%。以流式细胞术结果为基准,自动化 KB 计数的结果比手动计数更准确,平均偏差更小,一致性界限更窄。自动化 KB 计数的精密度高于手动 KB 计数(组内相关系数 0.996 比 0.79)。自动化 KB 计数的效率比经认证的技术员手动计数高 200 倍。
自动化 KB 计数可提供准确、精确且高效的 FMH 定量结果。