Comparison of parameter agreement for characterization of corneal subbasal nerve plexus in the whorl-like region and central cornea using in vivo confocal microscopy.

机构信息

Tianjin Medical University Eye Hospital, Eye Institute and School of Optometry, No. 251 Fukang Road, Xiqing District, Tianjin, 300384, China.

Shanxi Eye Hospital, Taiyuan, China.

出版信息

Int Ophthalmol. 2021 Feb;41(2):559-565. doi: 10.1007/s10792-020-01607-x. Epub 2020 Nov 9.

Abstract

PURPOSE

To compare the reliability of the whorl-like region with that of the central cornea for accurate assessment of corneal subbasal nerve plexus (SNP) by analyzing the parameter variability of these two anatomical regions in repeated measurements.

METHODS

Participants were scanned in the central cornea and whorl-like region with in vivo confocal microscopy on three occasions by two examiners within a time span of one week. Coefficients of repeatability (CoR), intra-class correlation coefficient (ICC), and Bland-Altman scatter plots with 95% limits of agreement (LOA) in the central cornea and whorl-like region were calculated, respectively, based on the nerve fiber length, then the inter-observer and intra-observer agreement were compared between these two anatomical regions.

RESULTS

The inter-observer ICC was 0.945, the inter-observer CoR was 0.052, the intra-observer ICC was 0.936, and the inter-observer CoR was 0.046, with narrow 95% LOA within 1 standard deviation in the whorl-like region, whereas the inter-observer ICC was 0.600, the inter-observer CoR was 0.207, the intra-observer ICC was 0.206, and the intra-observer CoR was 0.253, with 95% LOA nearly threefold wider than the standard deviation in the central cornea.

CONCLUSIONS

Nerve parameter in the whorl-like region showed higher inter-observer and intra-observer agreement than that of the central cornea. The whorl-like region is a more reliable site for accurate assessment of SNP.

摘要

目的

通过分析这两个解剖区域在重复测量中的参数变异性,比较涡旋状区域与中央角膜在准确评估角膜基底神经丛(SNP)方面的可靠性。

方法

在一周内,由两名检查者在活体共聚焦显微镜下对参与者的中央角膜和涡旋状区域进行三次扫描。分别计算中央角膜和涡旋状区域的神经纤维长度的重复性系数(CoR)、组内相关系数(ICC)和 Bland-Altman 散点图及其 95%一致性界限(LOA),并比较这两个解剖区域的观察者间和观察者内一致性。

结果

观察者间 ICC 为 0.945,观察者间 CoR 为 0.052,观察者内 ICC 为 0.936,观察者内 CoR 为 0.046,涡旋状区域的 95% LOA 在 1 个标准差内较窄,而中央角膜的观察者间 ICC 为 0.600,观察者间 CoR 为 0.207,观察者内 ICC 为 0.206,观察者内 CoR 为 0.253,95% LOA 几乎是标准差的三倍宽。

结论

涡旋状区域的神经参数显示出比中央角膜更高的观察者间和观察者内一致性。涡旋状区域是准确评估 SNP 的更可靠部位。