Imaging of melanin in the eye is important as the melanin is structurally associated with some ocular diseases, such as age-related macular degeneration. Although optical coherence tomography (OCT) cannot distinguish tissues containing the melanin from other tissues intrinsically, polarization-sensitive OCT (PS-OCT) can detect the melanin through spatial depolarization of the backscattered light from the melanin granules. Entropy is one of the depolarization metrics that can be used to detect malanin granules in PS-OCT and valuable quantitative information on ocular tissue abnormalities can be retrived by correlating entropy with the melanin concentration. In this study, we investigate a relationship between the melanin concentration and some depolarization metrics including the entropy, and show that the entropy is linearly proportional to the melanin concentration in double logarithmic scale when noise bias is corrected for the entropy. In addition, we also confirm that the entropy does not depend on the incident state of polarization using the experimental data, which is one of important attributes that depolarization metrics should have. The dependence on the incident state of polarization is also analyzed for other depolarization metrics.
眼内黑色素的成像很重要,因为黑色素与一些眼部疾病(如年龄相关性黄斑变性)在结构上有关。虽然光学相干断层扫描(OCT)不能从本质上区分含有黑色素的组织和其他组织,但偏振敏感 OCT(PS-OCT)可以通过从黑色素颗粒向后散射的光的空间去偏振来检测黑色素。熵是可用于检测 PS-OCT 中黑色素颗粒的去偏度量之一,通过将熵与黑色素浓度相关联,可以获得有关眼部组织异常的有价值的定量信息。在这项研究中,我们研究了黑色素浓度与一些去偏度量(包括熵)之间的关系,并表明当对熵进行噪声偏差校正时,在双对数标度下,熵与黑色素浓度呈线性正比关系。此外,我们还使用实验数据证实了熵不依赖于偏振的入射状态,这是去偏度量应具有的重要属性之一。还对其他去偏度量进行了对偏振入射状态的依赖性分析。