Suppr超能文献

一项关于核医学诊疗中放射性核素校准仪准确性的国际多中心调查。

An international multi-center investigation on the accuracy of radionuclide calibrators in nuclear medicine theragnostics.

作者信息

Saldarriaga Vargas Clarita, Bauwens Matthias, Pooters Ivo N A, Pommé Stefaan, Peters Steffie M B, Segbers Marcel, Jentzen Walter, Vogg Andreas, van Velden Floris H P, Meyer Viol Sebastiaan L, Gotthardt Martin, Mottaghy Felix M, Wildberger Joachim E, Covens Peter, Wierts Roel

机构信息

Radiation Protection Dosimetry and Calibrations, Belgian Nuclear Research Centre (SCK CEN), Mol, Belgium.

In vivo Cellular and Molecular Imaging, Vrije Universiteit Brussel, Jette, Belgium.

出版信息

EJNMMI Phys. 2020 Nov 23;7(1):69. doi: 10.1186/s40658-020-00338-3.

Abstract

BACKGROUND

Personalized molecular radiotherapy based on theragnostics requires accurate quantification of the amount of radiopharmaceutical activity administered to patients both in diagnostic and therapeutic applications. This international multi-center study aims to investigate the clinical measurement accuracy of radionuclide calibrators for 7 radionuclides used in theragnostics: Tc, In, I, I, I, Lu, and Y.

METHODS

In total, 32 radionuclide calibrators from 8 hospitals located in the Netherlands, Belgium, and Germany were tested. For each radionuclide, a set of four samples comprising two clinical containers (10-mL glass vial and 3-mL syringe) with two filling volumes were measured. The reference value of each sample was determined by two certified radioactivity calibration centers (SCK CEN and JRC) using two secondary standard ionization chambers. The deviation in measured activity with respect to the reference value was determined for each radionuclide and each measurement geometry. In addition, the combined systematic deviation of activity measurements in a theragnostic setting was evaluated for 5 clinically relevant theragnostic pairs: I/I, I/I, Lu/In, Y/Tc, and Y/In.

RESULTS

For Tc, I, and Lu, a small minority of measurements were not within ± 5% range from the reference activity (percentage of measurements not within range: Tc, 6%; I, 14%; Lu, 24%) and almost none were outside ± 10% range. However, for In, I, I, and Y, more than half of all measurements were not accurate within ± 5% range (In, 51%; I, 83%; I, 63%; Y, 61%) and not all were within ± 10% margin (In, 22%; I, 35%; I, 15%; Y, 25%). A large variability in measurement accuracy was observed between radionuclide calibrator systems, type of sample container (vial vs syringe), and source-geometry calibration/correction settings used. Consequently, we observed large combined deviations (percentage deviation > ± 10%) for the investigated theragnostic pairs, in particular for Y/In, I/I, and Y/Tc.

CONCLUSIONS

Our study shows that substantial over- or underestimation of therapeutic patient doses is likely to occur in a theragnostic setting due to errors in the assessment of radioactivity with radionuclide calibrators. These findings underline the importance of thorough validation of radionuclide calibrator systems for each clinically relevant radionuclide and sample geometry.

摘要

背景

基于诊疗一体化的个性化分子放射治疗需要在诊断和治疗应用中准确量化给予患者的放射性药物活度。这项国际多中心研究旨在调查用于诊疗一体化的7种放射性核素(锝、铟、碘、碘、碘、镥和钇)的放射性核素校准仪的临床测量准确性。

方法

总共测试了来自荷兰、比利时和德国8家医院的32台放射性核素校准仪。对于每种放射性核素,测量一组包含两个临床容器(10毫升玻璃瓶和3毫升注射器)、两种填充体积的四个样品。每个样品的参考值由两个经认证的放射性校准中心(SCK CEN和JRC)使用两个二级标准电离室确定。确定每种放射性核素和每种测量几何形状下测量活度相对于参考值的偏差。此外,还评估了诊疗一体化环境中5种临床相关诊疗对(碘/碘、碘/碘、镥/铟、钇/锝和钇/铟)活度测量的综合系统偏差。

结果

对于锝、碘和镥,少数测量值不在参考活度的±5%范围内(不在范围内的测量百分比:锝,6%;碘,14%;镥,24%),几乎没有超出±10%范围的。然而,对于铟、碘、碘和钇,超过一半的测量值在±5%范围内不准确(铟,51%;碘,83%;碘,63%;钇,61%),并非所有测量值都在±10%范围内(铟,22%;碘,35%;碘,15%;钇,25%)。在放射性核素校准仪系统、样品容器类型(玻璃瓶与注射器)以及使用的源几何校准/校正设置之间,观察到测量准确性存在很大差异。因此,我们观察到所研究的诊疗对存在较大的综合偏差(百分比偏差>±10%),特别是对于钇/铟、碘/碘和钇/锝。

结论

我们的研究表明,在诊疗一体化环境中,由于放射性核素校准仪对放射性评估的误差,很可能会出现对治疗患者剂量的严重高估或低估。这些发现强调了针对每种临床相关放射性核素和样品几何形状对放射性核素校准仪系统进行全面验证的重要性。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/251f/7683758/ff22a0dbc783/40658_2020_338_Fig1_HTML.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验