Foot Ankle Surg. 2021 Dec;27(8):855-859. doi: 10.1016/j.fas.2020.11.003. Epub 2020 Nov 25.
The medial distal tibial angle (MDTA) is used for measurement of ankle alignment. Standard to measure MDTA is weightbearing mortise view. EOS imaging becomes more popular for limb alignment analysis using low-dose radiation. As MDTA might vary in EOS, comparison between both radiographic techniques has been performed.
MDTA was compared between both techniques in 43 cases by defining the mechanical tibial axis in different ways (X-ray low, EOS low, EOS high). For each method MDTA, intra- and interobserver reliability has been compared.
The correlation between the different methods were measured by ICC (intraclass coefficient) and were ICC 0.86 (X-ray low/EOS low), ICC 0.85 (X-ray low/EOS high) and ICC 0.97 (EOS low/EOS high). Intra- and interobserver reliability were in each case ICC > 0.95.
ICC showed a substantial to excellent agreement between all methods. EOS is appropriate to determine MDTA and can be used for assessment of coronar deformities of the distal tibia.