Suppr超能文献

NIDEK OPD-Scan III 在近视儿童中瞳孔大小测量的可重复性。

Repeatability of pupil size measurements with NIDEK OPD-Scan III in myopic children.

机构信息

School of Optometry, The Hong Kong Polytechnic University, Hong Kong SAR, China.

Department of Ophthalmology, The University of Hong Kong, Hong Kong SAR, China.

出版信息

Ophthalmic Physiol Opt. 2021 Mar;41(2):431-436. doi: 10.1111/opo.12774. Epub 2020 Dec 8.

Abstract

OBJECTIVE

To investigate the repeatability of pupil size measurements, determined from the NIDEK OPD-Scan III, of myopic eyes in children wearing single-vision spectacles (SVS), undergoing orthokeratology (OK) and receiving combined treatment of 0.01% atropine and orthokeratology (AOK).

METHODS

A single examiner took two sets of mesopic pupil size measurements from the right eyes of 80 children (SVS:16, OK: 34, AOK: 30) (mesopic 1 and mesopic 2) and two sets of photopic (photopic 1 and photopic 2) pupil size measurements (internal light source) using the NIDEK OPD-Scan III. Each set was taken at a single visit, 15 min apart. Subjects in the OK and AOK groups had undergone treatment for at least 6 months, and the SVS subjects had no previous myopia treatment other than wearing spectacles. Repeatability of measurements was analysed using intraclass correlation coefficient (ICC), Bland-Altman (BA) plot, and coefficient of repeatability (CoR).

RESULTS

Mesopic 1 and 2 values were not significantly different among the three groups (p = 0.56, 0.77), hence the data were pooled for each set of measurements (mean [±S.D.] values were 7.09 ± 0.61 and 7.12 ± 0.61 mm, respectively) for further analyses. ICC for repeated mesopic pupil size measurements was 0.98 (95% CI, 0.98-0.99). No significant correlation was found between the means of the mesopic 1 and 2 measurements and their differences (Pearson's r = -0.02, p = 0.83). A BA plot also demonstrated narrow 95% limits of agreement, with a CoR of 0.28 mm. Mean photopic 1 and 2 measurements for the AOK group (3.74 ± 0.46; 3.73 ± 0.43 mm, respectively) were significantly larger (p = 0.01; 0.009) than those of the SVS and OK groups, but no significant difference was found between the latter two groups (p > 0.05). Hence, photopic 1 and 2 measurements for the SVS and OK groups were pooled (SVS-OK) for further analyses. ICC for repeatability of the photopic measurements was 0.98 (95% CI, 0.96 to 0.99) for the SVS-OK and AOK groups. The differences between photopic 1 and 2 measurements were not significantly correlated with their means, either in the AOK group (Pearson's r = -0.25, p = 0.19), or in the pooled SVS-OK group (Pearson's r = -0.04, p = 0.78). BA plots also showed narrow 95% limits of agreement, and CoR was 0.25 mm and 0.23 mm in the AOK and pooled SVS-OK groups, respectively.

CONCLUSION

Mesopic and photopic pupil size measurement using the NIDEK OPD-Scan III was highly repeatable and suitable for use on children.

摘要

目的

研究使用尼德克 OPD-Scan III 对配戴单光眼镜(SVS)、接受角膜塑形术(OK)和接受 0.01%阿托品联合角膜塑形术(AOK)的近视儿童的瞳孔大小测量值的可重复性。

方法

一名检查者对 80 名儿童(SVS:16 名,OK:34 名,AOK:30 名)(暗视 1 和暗视 2)右眼的两组中值瞳孔大小(暗视 1 和暗视 2)和两组明视瞳孔大小(明视 1 和明视 2)(内部光源)进行测量。每组在一次就诊时相隔 15 分钟进行两次测量。OK 和 AOK 组的受试者接受了至少 6 个月的治疗,SVS 组除了戴眼镜外,没有进行过其他近视治疗。使用组内相关系数(ICC)、Bland-Altman(BA)图和可重复性系数(CoR)分析测量的可重复性。

结果

三组之间暗视 1 和 2 值无显著性差异(p=0.56,0.77),因此将每组测量数据合并(平均值[±标准差]分别为 7.09±0.61 和 7.12±0.61mm)进行进一步分析。暗视瞳孔大小的 ICC 为 0.98(95%可信区间,0.98-0.99)。暗视 1 和 2 测量值的平均值及其差异之间无显著相关性(Pearson r=-0.02,p=0.83)。BA 图也显示了狭窄的 95%一致性界限,CoR 为 0.28mm。AOK 组的明视 1 和 2 测量值(3.74±0.46;3.73±0.43mm)明显大于 SVS 和 OK 组(p=0.01;0.009),但后两组之间无显著差异(p>0.05)。因此,SVS 和 OK 组的明视 1 和 2 测量值被合并(SVS-OK)进行进一步分析。SVS-OK 和 AOK 组的明视测量值的 ICC 为 0.98(95%可信区间,0.96 至 0.99)。AOK 组(Pearson r=-0.25,p=0.19)或合并的 SVS-OK 组(Pearson r=-0.04,p=0.78)中,明视 1 和 2 测量值的差异与其平均值均无显著相关性。BA 图也显示了狭窄的 95%一致性界限,AOK 组和合并的 SVS-OK 组的 CoR 分别为 0.25mm 和 0.23mm。

结论

使用尼德克 OPD-Scan III 进行中值和明视瞳孔大小测量具有高度可重复性,适用于儿童。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验