School of Optometry, The Hong Kong Polytechnic University, Hong Kong SAR, China.
Department of Ophthalmology, The University of Hong Kong, Hong Kong SAR, China.
Ophthalmic Physiol Opt. 2021 Mar;41(2):431-436. doi: 10.1111/opo.12774. Epub 2020 Dec 8.
To investigate the repeatability of pupil size measurements, determined from the NIDEK OPD-Scan III, of myopic eyes in children wearing single-vision spectacles (SVS), undergoing orthokeratology (OK) and receiving combined treatment of 0.01% atropine and orthokeratology (AOK).
A single examiner took two sets of mesopic pupil size measurements from the right eyes of 80 children (SVS:16, OK: 34, AOK: 30) (mesopic 1 and mesopic 2) and two sets of photopic (photopic 1 and photopic 2) pupil size measurements (internal light source) using the NIDEK OPD-Scan III. Each set was taken at a single visit, 15 min apart. Subjects in the OK and AOK groups had undergone treatment for at least 6 months, and the SVS subjects had no previous myopia treatment other than wearing spectacles. Repeatability of measurements was analysed using intraclass correlation coefficient (ICC), Bland-Altman (BA) plot, and coefficient of repeatability (CoR).
Mesopic 1 and 2 values were not significantly different among the three groups (p = 0.56, 0.77), hence the data were pooled for each set of measurements (mean [±S.D.] values were 7.09 ± 0.61 and 7.12 ± 0.61 mm, respectively) for further analyses. ICC for repeated mesopic pupil size measurements was 0.98 (95% CI, 0.98-0.99). No significant correlation was found between the means of the mesopic 1 and 2 measurements and their differences (Pearson's r = -0.02, p = 0.83). A BA plot also demonstrated narrow 95% limits of agreement, with a CoR of 0.28 mm. Mean photopic 1 and 2 measurements for the AOK group (3.74 ± 0.46; 3.73 ± 0.43 mm, respectively) were significantly larger (p = 0.01; 0.009) than those of the SVS and OK groups, but no significant difference was found between the latter two groups (p > 0.05). Hence, photopic 1 and 2 measurements for the SVS and OK groups were pooled (SVS-OK) for further analyses. ICC for repeatability of the photopic measurements was 0.98 (95% CI, 0.96 to 0.99) for the SVS-OK and AOK groups. The differences between photopic 1 and 2 measurements were not significantly correlated with their means, either in the AOK group (Pearson's r = -0.25, p = 0.19), or in the pooled SVS-OK group (Pearson's r = -0.04, p = 0.78). BA plots also showed narrow 95% limits of agreement, and CoR was 0.25 mm and 0.23 mm in the AOK and pooled SVS-OK groups, respectively.
Mesopic and photopic pupil size measurement using the NIDEK OPD-Scan III was highly repeatable and suitable for use on children.
研究使用尼德克 OPD-Scan III 对配戴单光眼镜(SVS)、接受角膜塑形术(OK)和接受 0.01%阿托品联合角膜塑形术(AOK)的近视儿童的瞳孔大小测量值的可重复性。
一名检查者对 80 名儿童(SVS:16 名,OK:34 名,AOK:30 名)(暗视 1 和暗视 2)右眼的两组中值瞳孔大小(暗视 1 和暗视 2)和两组明视瞳孔大小(明视 1 和明视 2)(内部光源)进行测量。每组在一次就诊时相隔 15 分钟进行两次测量。OK 和 AOK 组的受试者接受了至少 6 个月的治疗,SVS 组除了戴眼镜外,没有进行过其他近视治疗。使用组内相关系数(ICC)、Bland-Altman(BA)图和可重复性系数(CoR)分析测量的可重复性。
三组之间暗视 1 和 2 值无显著性差异(p=0.56,0.77),因此将每组测量数据合并(平均值[±标准差]分别为 7.09±0.61 和 7.12±0.61mm)进行进一步分析。暗视瞳孔大小的 ICC 为 0.98(95%可信区间,0.98-0.99)。暗视 1 和 2 测量值的平均值及其差异之间无显著相关性(Pearson r=-0.02,p=0.83)。BA 图也显示了狭窄的 95%一致性界限,CoR 为 0.28mm。AOK 组的明视 1 和 2 测量值(3.74±0.46;3.73±0.43mm)明显大于 SVS 和 OK 组(p=0.01;0.009),但后两组之间无显著差异(p>0.05)。因此,SVS 和 OK 组的明视 1 和 2 测量值被合并(SVS-OK)进行进一步分析。SVS-OK 和 AOK 组的明视测量值的 ICC 为 0.98(95%可信区间,0.96 至 0.99)。AOK 组(Pearson r=-0.25,p=0.19)或合并的 SVS-OK 组(Pearson r=-0.04,p=0.78)中,明视 1 和 2 测量值的差异与其平均值均无显著相关性。BA 图也显示了狭窄的 95%一致性界限,AOK 组和合并的 SVS-OK 组的 CoR 分别为 0.25mm 和 0.23mm。
使用尼德克 OPD-Scan III 进行中值和明视瞳孔大小测量具有高度可重复性,适用于儿童。