We describe systematic troubleshooting of the carry-over of neuropeptide Y (NPY) in LC-MS analysis. The objective was to remove candidate parts of the LC-MS system that are responsible for carry over one-by-one. The findings indicate that the carry-over of NPY occurs on the column, particularly in the guard column and at the consumable seals of the sample-needle and high-pressure valves. The methodology demonstrates that it is possible to troubleshoot carry-over in an LC-MS system in a systematic and logical manner.
我们描述了在液相色谱-质谱联用(LC-MS)分析中对神经肽Y(NPY)残留进行系统故障排查的过程。目的是逐一排查LC-MS系统中导致残留的可能部件。结果表明,NPY的残留发生在色谱柱上,尤其是在保护柱以及样品针和高压阀的易耗密封处。该方法表明,以系统且合乎逻辑的方式对LC-MS系统中的残留进行故障排查是可行的。