Department of Chemistry and Applied Biosciences, ETH Zurich, CH-8093, Zurich, Switzerland.
Rapid Commun Mass Spectrom. 2021 Mar 30;35(6):e9021. doi: 10.1002/rcm.9021.
Fragrances are organic compounds with pleasant odors that are widely used in every aspect of our daily life; some fragrance ingredients can cause allergic reactions. Hence, the qualitative and quantitative analysis of fragrance allergens can prevent consumers coming into contact with these compounds. In this study, we evaluated the ability of a dielectric barrier discharge ionization (DBDI) source for analyzing allergens that occur in fragrances.
A home-built liquid-infusion device was used to evaporate the liquid samples. An active capillary plasma ionization source, which is based on a dielectric barrier discharge, was used to ionize the analytes. Mass spectra were acquired in positive ion mode with an LTQ Orbitrap mass spectrometer.
Seven typical fragrance allergens were analyzed in this study. The limits of detections (LODs) were as low as 0.0001 ppm and a linear dynamic range of 2-3 orders of magnitude was achieved. Allergens in five different perfume products were successfully analyzed and quantified by this method, with analysis times of less than 1 min per sample.
This work introduces a DBDI-MS-based analytical method for detecting and quantifying fragrance allergens. Since DBDI has the advantages of high sensitivity, simple operation and fast analysis time, it is very suitable for the rapid analysis of trace allergens in fragrances, and could easily be used for quality control of consumer products in the cosmetics market.
香料是具有宜人气味的有机化合物,广泛应用于我们日常生活的各个方面;一些香料成分会引起过敏反应。因此,对香料过敏原进行定性和定量分析可以防止消费者接触这些化合物。在这项研究中,我们评估了介电屏蔽放电电离(DBDI)源分析香料中过敏原的能力。
使用自制的液体注入装置蒸发液体样品。基于介电屏蔽放电的活性毛细管等离子体电离源用于使分析物电离。采用 LTQ Orbitrap 质谱仪以正离子模式采集质谱。
本研究分析了七种典型的香料过敏原。检测限低至 0.0001ppm,线性动态范围达到 2-3 个数量级。该方法成功分析和定量了五种不同香水产品中的过敏原,每个样品的分析时间不到 1 分钟。
本工作介绍了一种基于 DBDI-MS 的分析方法,用于检测和定量香料过敏原。由于 DBDI 具有灵敏度高、操作简单、分析时间短等优点,非常适合快速分析香料中的痕量过敏原,并且可以轻松用于化妆品市场中消费品的质量控制。