Suppr超能文献

儿童群体中耳蜗植入物早期激活对电极阻抗的影响。

Effect of early activation of cochlear implant on electrode impedance in pediatric population.

机构信息

King Abdullah Ear Specialist Center (KAESC), College of Medicine, King Saud University, Riyadh, Saudi Arabia.

Research Department, MED-EL GmbH, Riyadh, Saudi Arabia.

出版信息

Int J Pediatr Otorhinolaryngol. 2021 Jan;140:110543. doi: 10.1016/j.ijporl.2020.110543. Epub 2020 Dec 2.

Abstract

OBJECTIVES

To assess the evolution of electrode impedance after the early fitting of audio processors (activation after one-day) and classical fitting (activation after one-month) over an up-to-one year after cochlear implant (CI) surgery.

METHODS

A retrospective cohort study on Fifty-two CI recipients divided into two groups. The study group included 24 recipients (40 ears) who underwent early fitting, whereas the control group contained 28 recipients (40 ears) who underwent classical fitting. The electrode impedance was recorded during the surgery, switch-on session and at one, three, six, and twelve-months after the surgery.

RESULTS

In the study group, electrode impedance values obtained intraoperatively and at switch-on and one, three, six, and twelve-months were 4.89, 3.69, 6.52, 6.24, 6.05, and 5.81 KΩ, respectively, and only the switch-on and one-month values were significantly different (p < 0.0001). In the control group, electrode impedance values obtained intraoperatively and at switch-on and one, three, six, and twelve-months were 4.71, 7.19, 6.40, 6.05, and 5.73 KΩ, respectively. Thus, the electrode impedance value at switch-on was 52.65% (p < 0.001) greater than it intraoperatively. For both groups, the electrode impedance value at twelve-months was significantly higher than the respective intraoperative values (study group: 18.6% higher, P = 0.04; control group: 21.65% higher, P = 0.0001).

CONCLUSION

Electrode impedance was significantly lower in the study group compared to the control group until one month after the surgery. However, the electrode impedance at twelve-months after the CI was similar in both groups.

摘要

目的

评估人工耳蜗植入术后 1 天(早期适配,即开机)和 1 个月(经典适配)后 1 年的电极阻抗变化。

方法

回顾性队列研究纳入 52 名人工耳蜗植入患者,分为两组。研究组(n=24,40 耳)行早期适配,对照组(n=28,40 耳)行经典适配。术中、开机时及术后 1、3、6、12 个月记录电极阻抗。

结果

研究组术中及开机时和术后 1、3、6、12 个月的电极阻抗值分别为 4.89、3.69、6.52、6.24、6.05 和 5.81 KΩ,仅开机时和术后 1 个月的阻抗值有显著差异(p<0.0001)。对照组术中及开机时和术后 1、3、6、12 个月的电极阻抗值分别为 4.71、7.19、6.40、6.05 和 5.73 KΩ,因此开机时的阻抗值比术中高 52.65%(p<0.001)。两组术后 12 个月的电极阻抗值均明显高于术中值(研究组:高 18.6%,P=0.04;对照组:高 21.65%,P=0.0001)。

结论

与对照组相比,研究组术后 1 个月内电极阻抗明显降低,但两组术后 12 个月的电极阻抗相似。