Tikhomirov I B, Turchaninova V F, Selivanenko V T, Staferov V A
Kosm Biol Aviakosm Med. 1977 Jul-Aug;11(4):70-4.
Experiments were carried out to explore the possible use of tetrapolar rheography in the assessment of pulse blood supply to a certain area and quantitative characteristics of cardiac output and blood flow in the limbs. Sixty five subjects were examined: fifty subjects had congenital diseases of the heart and the major vessels and 15 subjects were healthy. The volume-pulse blood supply was measured by means of a tetrapolar rheograph. Calculations were based on the Kedrov formula. Control measurements of cardiac output were made by the dye-dilution method (Evans blue). The correlation coefficient was 0.96. This gives evidence that the above instrument can be used to determine cardiac output. Studies of the pulse blood supply to the limbs show that the measurements may reflect not only marked pathology but also changes due to functional vascular shifts.
开展了多项实验,以探索四极血流图在评估某一区域的脉冲血液供应以及心输出量和肢体血流的定量特征方面的可能用途。对65名受试者进行了检查:50名受试者患有先天性心脏和大血管疾病,15名受试者健康。通过四极血流图测量容积脉冲血液供应。计算基于凯德罗夫公式。采用染料稀释法(伊文思蓝)进行心输出量的对照测量。相关系数为0.96。这证明上述仪器可用于测定心输出量。对肢体脉冲血液供应的研究表明,这些测量不仅可以反映明显的病理情况,还可以反映由于功能性血管变化而引起的改变。