The stroke blood volume and the MBV were determined by tetrapolar rheocardiography at rest and under the effect of muscular exertion in neurosurgical patients (35), in surgical patients (20) and in a control group of healthy subjects (10). A Soviet device, the Ppigamma2-02 tetrapolar rheocardiograph was employed. The method of tetrapolar rheocardiography is simple, atraumatic and provides for prolonged continuous recording. It is sufficiently precise as a method of indirect determination of the minute volume: comparison with data of the radioisotope method shows the precision of tetrapolar rheocardiography to be +/- 5.5%. Its main disadvantage consists in the difficult calculation of MBV in deep, frequent and irregular respiration when the rheocardiogram waves are sharply distorted.
通过四极心流变图法测定了神经外科患者(35例)、外科手术患者(20例)以及健康受试者对照组(10例)在静息状态和肌肉运动影响下的中风血容量和每搏量。使用的是苏联的一种设备,即Ppigamma2 - 02四极心流变仪。四极心流变图法操作简单、无创,可进行长时间连续记录。作为一种间接测定每分容量的方法,它足够精确:与放射性同位素法的数据比较表明,四极心流变图法的精确度为±5.5%。其主要缺点在于,当心动流变图波形严重失真时,在深度、频繁且不规则呼吸情况下,每搏量的计算较为困难。