Suppr超能文献

原发性先天性青光眼患者回弹式眼压计(Icare ic200)与压平式眼压计(Perkins)测量结果的一致性

Agreement between rebound (Icare ic200) and applanation tonometry (Perkins) in patients with primary congenital glaucoma.

作者信息

Morales-Fernandez Laura, Pérez-García Pilar, Saenz-Frances Federico, Molero-Senosiain Mercedes, Garcia-Saenz Sofia, Dora Mendez Carmen, Santos Bueso Enrique, Garcia-Feijoo Julian, Martinez-de-la-Casa Jose M

机构信息

Ophthalmology Unit, Hospital Clinico San Carlos, Department of Ophthalmology and ORL, Faculty of Medicine, University Complutense (Madrid), Institute "Investigacion Sanitaria del Hospital Clinico San Carlos" (IdISSC), Madrid, Spain.

Hospital Universitario Quiron Pozuelo, Madrid, Spain.

出版信息

Acta Ophthalmol. 2021 Sep;99(6):663-668. doi: 10.1111/aos.14701. Epub 2020 Dec 23.

Abstract

PURPOSE

To examine agreement between intraocular pressure (IOP) measurements made using the rebound tonometer Icare ic200 (RT200) and the Perkins handheld applanation tonometer (PAT) in patients with primary congenital glaucoma (PCG). The impacts of several covariables on measurements using the two devices were also assessed.

MATERIALS AND METHODS

Intraocular pressure measurements were made in a single session in 86 eyes of 86 patients with PCG (46 under anaesthesia, 40 in the office). The order was RT200 then PAT. The variables age, central corneal thickness (CCT), corneal state and anaesthesia were recorded in each patient. Data were compared by determining interclass correlation coefficients (ICC) for each tonometer and representing the differences detected as Bland-Altman plots. Effects of covariables were assessed through univariate and multivariate regression.

RESULTS

Mean IOP difference between tonometers (RT200 minus PAT) was 1.26 mmHg (95%: 0.22-2.31). Absolute agreement (ICC) was 0.73 (95% CI: 0.62-0.82). Lower and upper limits of agreement (95%) were -8.06 mmHg (95% CI: -9.87 to -6.25) and 10.59 mmHg (95% CI: 8.77-12.40), respectively. The tonometers showed systematic differences (a = -4.63 mmHg; 95% CI: -9.11 to -1.44) and proportional differences; for each mmHg increase in PAT-IOP, the RT200 reading increased by 1.28 mmHg (b = 1.28; 95% CI: 1.12-1.53). None of the variables tested as predictors were able to explain differences between the tonometers.

CONCLUSIONS

Despite the good overall agreement between both tonometers, caution should be taken in high values of IOP, considering the interchangeability of its readings as systematic and proportional differences appear to exist between both methods.

摘要

目的

研究使用回弹式眼压计Icare ic200(RT200)和珀金手持式压平眼压计(PAT)测量原发性先天性青光眼(PCG)患者眼压的一致性。还评估了几个协变量对使用这两种设备测量结果的影响。

材料与方法

对86例PCG患者的86只眼进行单次眼压测量(46例在麻醉下,40例在诊室)。测量顺序为先RT200后PAT。记录每位患者的年龄、中央角膜厚度(CCT)、角膜状态和麻醉情况。通过确定每种眼压计的组内相关系数(ICC)并以Bland-Altman图表示检测到的差异来比较数据。通过单变量和多变量回归评估协变量的影响。

结果

眼压计之间的平均眼压差异(RT200减去PAT)为1.26 mmHg(95%:0.22 - 2.31)。绝对一致性(ICC)为0.73(95%CI:0.62 - 0.82)。一致性的下限和上限(95%)分别为 - 8.06 mmHg(95%CI: - 9.87至 - 6.25)和10.59 mmHg(95%CI:8.77 - 12.40)。眼压计显示出系统差异(a = - 4.63 mmHg;95%CI: - 9.11至 - 1.44)和成比例差异;PAT眼压每升高1 mmHg,RT200读数升高1.28 mmHg(b = 1.28;95%CI:1.12 - 1.53)。作为预测指标测试的变量均无法解释眼压计之间的差异。

结论

尽管两种眼压计总体一致性良好,但考虑到两种方法之间似乎存在系统差异和成比例差异,其读数的互换性,对于高眼压值仍应谨慎。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验