成人正常角膜眼的 Icare ic200 眼压测量与 Goldmann 压平眼压计的一致性。

Agreement of Intraocular Pressure Measurement of Icare ic200 with Goldmann Applanation Tonometer in Adult Eyes with Normal Cornea.

机构信息

VST center for Glaucoma, L V Prasad Eye Institute, Hyderabad, India.

Narayana Nethralaya, Bannerghatta Road, Bangalore, India.

出版信息

Ophthalmol Glaucoma. 2021 Jan-Feb;4(1):89-94. doi: 10.1016/j.ogla.2020.08.004. Epub 2020 Aug 13.

Abstract

PURPOSE

To study the agreement between the Icare ic200 (ICare Finland Oy, Helsinki, Finland) and the Goldmann Applanation Tonometer (GAT) in the measurement of intraocular pressure (IOP) in adult eyes.

DESIGN

Noninterventional, cross-sectional study.

PARTICIPANTS

A total of 156 eyes of 156 adult participants with clear corneas were included.

METHODS

The IOP measurements were obtained with the Icare ic200 by 1 observer followed by GAT readings by a second masked observer. The central corneal thickness (CCT) and biometry of all subjects were recorded.

MAIN OUTCOME MEASURES

The agreement between Icare ic200 and GAT was measured using the Bland-Altman plot.

RESULTS

The mean age ± standard deviation of subjects was 55.3 ± 13.7 years. The GAT IOP ranged from 6 to 50 mmHg with a mean IOP of 19.5 ± 8.8 mmHg. The Icare ic200 IOP ranged from 7.4 to 50 mmHg with a mean IOP of 20.8 ± 9.3 mmHg. The mean difference between the IOP measurement of GAT and Icare ic200 was -1.27 mmHg with the 95% limits of agreement (LoA) ranging from -3.4 to 0.9 mmHg for all ranges of IOP. The mean difference (95% LoA) between the IOP measurement of GAT and Icare ic200 was -1 mmHg (-3 to 1 mmHg) and -1.8 mmHg (-4 to 0.2 mmHg) for a GAT IOP ≤21 mmHg and >21 mmHg, respectively. The CCT, axial length, age, and gender did not significantly affect the difference in measurement of IOP between the 2 tonometers. However, for every 1-mmHg increase in GAT IOP, the difference between the 2 tonometers increased by 0.04 mmHg (P < 0.001).

CONCLUSIONS

In our study, the Icare ic200 overestimated the IOP. The overestimation increased as the baseline IOP increased. The agreement between the IOP measurement by GAT and Icare ic200 was <2 mmHg at all ranges of IOP. The narrow LoA between the tonometers for an IOP <21 mmHg makes it a useful alternative to GAT in this pressure range.

摘要

目的

研究 Icare ic200(芬兰埃卡利公司,赫尔辛基,芬兰)与 Goldmann 压平眼压计(GAT)在成人眼内眼压(IOP)测量中的一致性。

设计

非干预性、横断面研究。

参与者

共纳入 156 例 156 只成年受试者,其角膜透明。

方法

由 1 名观察者使用 Icare ic200 进行 IOP 测量,然后由第 2 名掩蔽观察者进行 GAT 读数。记录所有受试者的中央角膜厚度(CCT)和生物测量值。

主要观察指标

采用 Bland-Altman 图比较 Icare ic200 和 GAT 的一致性。

结果

受试者的平均年龄(标准差)为 55.3±13.7 岁。GAT 的 IOP 范围为 6 至 50mmHg,平均 IOP 为 19.5±8.8mmHg。Icare ic200 的 IOP 范围为 7.4 至 50mmHg,平均 IOP 为 20.8±9.3mmHg。GAT 和 Icare ic200 的 IOP 测量值之间的平均差值为-1.27mmHg,95%一致性区间(LoA)为-3.4 至 0.9mmHg,适用于所有 IOP 范围。GAT 和 Icare ic200 的 IOP 测量值之间的平均差值(95%LoA)分别为-1mmHg(-3 至 1mmHg)和-1.8mmHg(-4 至 0.2mmHg),GAT IOP 分别为≤21mmHg 和>21mmHg。CCT、眼轴长度、年龄和性别均未显著影响两种眼压计测量的 IOP 差值。然而,GAT IOP 每增加 1mmHg,两种眼压计之间的差值增加 0.04mmHg(P<0.001)。

结论

在本研究中,Icare ic200 高估了 IOP。随着基础 IOP 的升高,高估程度增加。GAT 和 Icare ic200 测量的 IOP 在所有 IOP 范围内的一致性均<2mmHg。两种眼压计在 IOP<21mmHg 时的 LoA 较窄,使其成为该压力范围内 GAT 的有用替代方法。