应用即时凝血监测仪的黏弹性指标建立犬的正常参考区间,并与血栓弹力描记术进行比较。

Establishment of normal reference intervals in dogs using a viscoelastic point-of-care coagulation monitor and its comparison with thromboelastography.

机构信息

Department of Clinical Sciences and Advanced Medicine, Veterinary Hospital of the University of Pennsylvania, Philadelphia, PA, USA.

出版信息

Vet Clin Pathol. 2020 Dec;49(4):567-573. doi: 10.1111/vcp.12926. Epub 2020 Dec 27.

Abstract

BACKGROUND

Viscoelastic coagulation devices are a useful adjunct to the evaluation of hemostasis in veterinary patients. VCM Vet is a point-of-care device that is simple in operation and could be used to diagnose and trend hemostatic abnormalities in sick patients. VCM Vet does not use activators.

OBJECTIVES

We aimed to establish reference intervals (RIs) for VCM Vet in a healthy adult canine population and concurrently perform thromboelastographic (TEG) analysis on these samples with and without tissue factor (TF) activation for RI comparisons.

METHODS

Duplicate VCM Vet tests were performed immediately upon sample collection. Two concurrent TEG tests were performed on the remaining blood, one citrated, untreated (CU), and one activated with TF at a 1:3600 dilution.

RESULTS

Fifty-two dogs were enrolled in the study. The following RIs were generated for VCM Vet machine 1 and 2, respectively: clot time (CT) (seconds) 163-480 and 172-457; clot formation time (CFT) (seconds) 104-288 and 94-252, α-angle (degrees) 41-65 and 44-66, and maximum clot firmness (MCF) (no units) 27-43 and 30-46. Moderate to good correlations were observed between the two machines with Lin's concordance correlation coefficients of 0.51-0.9 and a P < 0.002. TEG RIs were similar to previously reported values.

CONCLUSIONS

VCM Vet RIs were generated. Each VCM Vet device should have a unique RI established due to inter-device variability. Direct correlations of VCM Vet values with TEG parameters were not performed due to the narrow range of the normal values and the need to evaluate patients with a wide range of hemostatic abnormalities.

摘要

背景

黏弹性凝血设备是评估兽医患者止血功能的有用辅助手段。VCM Vet 是一种操作简单的即时检测设备,可用于诊断和监测患病动物的止血异常。VCM Vet 不使用激活剂。

目的

我们旨在为健康成年犬建立 VCM Vet 的参考区间(RI),并同时对这些样本进行血栓弹性描记术(TEG)分析,包括未激活和激活组织因子(TF)的情况下的 RI 比较。

方法

在采集样本后立即进行两次重复的 VCM Vet 检测。将剩余的血液分为两份,一份枸橼酸盐抗凝、未处理(CU),一份用 TF 以 1:3600 的稀释度激活,每份血液均进行两次同时的 TEG 检测。

结果

本研究共纳入 52 只犬。分别为 VCM Vet 机器 1 和 2 生成以下 RI:凝血时间(CT)(秒)163-480 和 172-457;凝血形成时间(CFT)(秒)104-288 和 94-252,α角(度)41-65 和 44-66,最大凝血硬度(MCF)(无单位)27-43 和 30-46。两台机器之间观察到中度至高度相关性,Lin 一致性相关系数为 0.51-0.9,P<0.002。TEG RI 与之前报道的值相似。

结论

生成了 VCM Vet 的 RI。由于设备间存在差异,每个 VCM Vet 设备都应建立独特的 RI。由于正常范围较窄,并且需要评估具有广泛止血异常的患者,因此未对 VCM Vet 值与 TEG 参数进行直接相关性分析。