Establishment of a reference interval for a novel viscoelastic coagulometer and comparison with thromboelastography in healthy cats.

机构信息

William R. Pritchard Veterinary Medical Teaching Hospital, School of Veterinary Medicine, University of California, Davis, Davis, USA.

Department of Surgical and Radiological Sciences, School of Veterinary Medicine, University of California, Davis, Davis, USA.

出版信息

Vet Clin Pathol. 2020 Dec;49(4):660-664. doi: 10.1111/vcp.12916. Epub 2020 Nov 18.

Abstract

BACKGROUND

Viscoelastic analysis provides information on the dynamics and strength of clot formation as well as clot stability. A novel point-of-care viscoelastic test (Viscoelastic Coagulation Monitor Vet, VCM Vet) could be more cost-effective, simpler to use, and more portable than thromboelastography (TEG).

OBJECTIVES

The primary aim of this study was to establish a feline reference interval (RI) for the VCM Vet. A secondary aim was to compare VCM Vet analysis with TEG in healthy cats.

METHODS

Fifty-six healthy cats were enrolled in this study. Linear regression was completed to determine whether age and CBC parameters were associated with the VCM Vet parameters and if TEG parameters were correlated with VCM Vet data. Statistical significance was set at P < .05.

RESULTS

Fifty-three VCM Vet tracings were used to determine RIs for healthy cats. The determined RIs were: clot time (CT) 104-438 seconds; clot formation time (CFT) 104.5-488 seconds; alpha angle (AA) 30.5°-70°; a10 13.8-32.7 VCM units; a20 19.2-40.1 VCM units; maximum clot formation (MCF) 22.5-44.8 VCM units; Lysis Index 30 (Li30) 92.9%-100.9%; and Lysis Index 45 (Li45) 92%-100%. Linear regression identified a strong positive correlation between the CT and R-time measured using the VCM Vet and TEG methods, respectively; no other parameters were correlated.

CONCLUSIONS

The use of VCM Vet is feasible in cats, and we determined the first described feline RIs for this test. In general, the VCM Vet data did not correlate with TEG in healthy cats.

摘要

背景

黏弹性分析可提供关于血栓形成的动力学和强度以及血栓稳定性的信息。一种新型的即时黏弹性测试(Vet 型黏弹性凝血监测仪,VCM Vet)可能比血栓弹性描记术(TEG)更具成本效益、使用更简单且更便于携带。

目的

本研究的主要目的是建立 VCM Vet 的猫科动物参考区间(RI)。次要目的是比较 VCM Vet 分析与健康猫的 TEG。

方法

本研究纳入了 56 只健康猫。进行线性回归以确定年龄和全血细胞计数参数是否与 VCM Vet 参数相关,以及 TEG 参数是否与 VCM Vet 数据相关。统计学意义设定为 P <.05。

结果

使用 53 个 VCM Vet 轨迹确定了健康猫的 RI。确定的 RI 为:凝血时间(CT)104-438 秒;凝血形成时间(CFT)104.5-488 秒;α角(AA)30.5°-70°;a10 为 13.8-32.7 VCM 单位;a20 为 19.2-40.1 VCM 单位;最大凝血形成(MCF)为 22.5-44.8 VCM 单位;Lysis Index 30(Li30)为 92.9%-100.9%;Lysis Index 45(Li45)为 92%-100%。线性回归确定了 VCM Vet 测量的 CT 和 R 时间与 TEG 方法分别测量的 CT 和 R 时间之间存在强烈的正相关;没有其他参数相关。

结论

VCM Vet 在猫中使用是可行的,我们确定了该测试的第一个描述的猫科动物 RI。一般来说,在健康猫中,VCM Vet 数据与 TEG 不相关。