Appl Spectrosc. 2021 Jul;75(7):867-881. doi: 10.1177/0003702820987601. Epub 2021 Feb 4.
Raman spectroscopy has been used extensively to calculate CO fluid density in many geological environments, based on the measurement of the Fermi diad split (Δ; cm) in the CO spectrum. While recent research has allowed the calibration of several Raman CO densimeters, there is a limit to the interlaboratory application of published equations. These calculate two classes of density values for the same measured Δ, with a deviation of 0.09 ± 0.02 g/cm on average. To elucidate the influence of experimental parameters on the calibration of Raman CO densimeters, we propose a bottom-up approach beginning with the calibration of a new equation, to evaluate a possible instrument-dependent variability induced by experimental conditions. Then, we develop bootstrapped confidence intervals for density estimate of existing equations to move the statistical analysis from a sample-specific to a population level. We find that Raman densimeter equations calibrated based on spectra acquired with similar spectral resolution calculate CO density values lying within standard errors of equations and are suitable for the interlaboratory application. The statistical analysis confirms that equations calibrated at similar spectral resolution calculate CO densities equivalent at 95% confidence, and each Raman densimeter does have a limit of applicability, statistically defined by a minimum Δ value, below which the error in calculated CO densities is too high.
拉曼光谱已广泛应用于许多地质环境中,通过测量 CO 光谱中的费米双简并分裂(Δ;cm)来计算 CO 流体密度。尽管最近的研究允许对几种拉曼 CO 密度计进行校准,但发表的方程在实验室间的应用存在限制。这些方程为相同的测量Δ计算了两类密度值,平均偏差为 0.09±0.02 g/cm。为了阐明实验参数对拉曼 CO 密度计校准的影响,我们提出了一种自下而上的方法,从校准新方程开始,以评估实验条件引起的可能与仪器有关的可变性。然后,我们为现有方程的密度估计值开发了自举置信区间,将统计分析从样本特定水平转移到总体水平。我们发现,基于具有相似光谱分辨率的光谱校准的拉曼密度计方程计算的 CO 密度值在方程的标准误差范围内,适用于实验室间的应用。统计分析证实,在相似光谱分辨率下校准的方程计算出的 CO 密度在 95%置信度下等效,并且每个拉曼密度计确实有一个适用范围的限制,这是由一个最低的 Δ 值来统计定义的,低于该值,计算出的 CO 密度的误差就太高了。