Olsen T
Department of Ophthalmology, University of Aarhus, Aarhus Kommunehospital, Denmark.
J Cataract Refract Surg. 1988 Jan;14(1):64-7. doi: 10.1016/s0886-3350(88)80066-x.
A method to measure the power of an intraocular lens (IOL) in situ using an extended range keratometer and a depth measuring apparatus is described. With this method, the IOL power was measured in ten patients with an anterior chamber IOL and in ten patients with a posterior chamber IOL and compared with the labeled power. The total mean error (measured power minus labeled power) was found to be 0.01 diopters (D) with a standard deviation of 0.22 D. The largest error was 0.40 D.