使用不同方法进行的牙髓长度测量:一项研究。

Endodontic Length Measurements Using Different Modalities: An Study.

作者信息

Nguyen Phuc Ngoc, Pham Khoa Van

机构信息

Department of Operative Dentistry and Endodontics, Faculty of Odonto-Stomatology, University of Medicine and Pharmacy at Ho Chi Minh City, Ho Chi Minh City, Vietnam.

出版信息

J Int Soc Prev Community Dent. 2020 Nov 24;10(6):752-758. doi: 10.4103/jispcd.JISPCD_357_20. eCollection 2020 Nov-Dec.

Abstract

INTRODUCTION

The aim of this study was to investigate and compare the accuracy of the 3D Endo software, conventional CBCT software Romexis Viewer, and the EAL E-Pex Pro in WL determination.

MATERIALS AND METHODS

Three hundred and two root canals in 110 intact human extracted molars were accessed. The actual lengths (ALs) were measured. Root canal lengths were measured using 3D Endo with proposed length (3D-PL) by software, correct length (3D-CL), Romexis Viewer, and the E-Pex Pro. The percentages of the measurements in the range of ±0.5 mm to the AL were compared using the Fisher's exact test. The paired test and Bland-Altman plots were calculated to detect the agreement of the four methods with the AL measurements. The statistical significance was set at 0.05.

RESULTS

The accuracies in the range of ±0.5 mm to the AL were 83.8%, 86.7%, 48.3%, and 99.7% for 3D-PL, 3D-CL, Romexis Viewer, and E-PexPro, respectively. There were agreements between 3D-PL and Romexis Viewer with the AL measurements.

CONCLUSION

The CBCT measurements using 3D Endo with the proposed length by the software and Romexis Viewer with the voxel size of 0.15 mm agreed with the AL measurements of the root canals.

摘要

引言

本研究的目的是调查并比较3D Endo软件、传统CBCT软件Romexis Viewer和EAL E-Pex Pro在工作长度测定方面的准确性。

材料与方法

对110颗完整的人类拔除磨牙中的302个根管进行通路预备。测量实际长度(AL)。使用3D Endo软件根据建议长度(3D-PL)、正确长度(3D-CL)、Romexis Viewer和E-Pex Pro测量根管长度。使用Fisher精确检验比较在AL的±0.5毫米范围内测量值的百分比。计算配对t检验和Bland-Altman图以检测这四种方法与AL测量值的一致性。设定统计学显著性水平为0.05。

结果

3D-PL、3D-CL、Romexis Viewer和E-Pex Pro在AL的±0.5毫米范围内的准确率分别为83.8%、86.7%、48.3%和99.7%。3D-PL和Romexis Viewer与AL测量值之间存在一致性。

结论

使用软件建议长度的3D Endo和体素大小为0.15毫米的Romexis Viewer进行的CBCT测量与根管的AL测量值一致。