Physics and Biomedical Engineering, Centre Hospitalier de l'Université de Montréal, Montréal, Canada.
Biomed Phys Eng Express. 2020 Feb 24;6(2):027001. doi: 10.1088/2057-1976/ab7500.
Quantitative SPECT studies require specific information about the equipment being used. Particularly in the context of therapeutic studies, the effect of dead-time can be significant and must be quantified. We explored different techniques for measuring the dead-time constant and applying dead-time corrections to the data.
The dead-time constant was measured on four similar SPECT/CT systems by following the response of the system to a uniform phantom initially containing 17 GBq of Lu-177 over a period of 23 days. It was then calculated using the two-source method with 1 332 MBq of Tc-99 m. The dead-time constant found was used to correct SPECT/CT phantom images either applying the correction by projection or globally on the image.
Both methods of calculating the dead-time constant produced equivalent results. However, the dead-time constant varied by as much as 8% between machines of the same model and manufacturer. Correcting for dead-time by projection rather than globally produced slightly more precise results (0.94% error rather than 2.59% error). The benefit of this correction technique will be dependent on the level of asymmetry in the patient as well as the magnitude of the dead-time correction effect.
quantification of the dead-time of a system can be performed quickly using the two-source method and any radioisotope. However, it is important to perform this measurement on every system being used. In vastly asymmetric images with high dead-time correction, correcting for dead-time by projection can be pertinent, increasing the precision of dosimetry calculations by several percent. However this additional gain may be within the error of SUV measurements for many clinical acquisitions.
定量 SPECT 研究需要有关所使用设备的特定信息。特别是在治疗研究的背景下,死时间的影响可能很大,必须进行量化。我们探讨了测量死时间常数和对数据进行死时间校正的不同技术。
通过在 23 天的时间内跟踪系统对最初包含 17GBq Lu-177 的均匀体模的响应,在四个类似的 SPECT/CT 系统上测量死时间常数。然后使用 1332MBq Tc-99m 的双源方法进行计算。使用找到的死时间常数通过投影或全局方式校正 SPECT/CT 体模图像。
计算死时间常数的两种方法均产生等效结果。然而,同一型号和制造商的机器之间,死时间常数的差异高达 8%。通过投影而不是全局校正死时间会产生稍微更精确的结果(0.94%的误差而不是 2.59%的误差)。这种校正技术的好处将取决于患者的不对称程度以及死时间校正效果的大小。
可以使用双源方法和任何放射性同位素快速量化系统的死时间。但是,必须对正在使用的每个系统执行此测量。在具有高死时间校正的极大不对称图像中,通过投影校正死时间可能很重要,通过几个百分点提高剂量计算的精度。但是,对于许多临床采集,此额外增益可能在 SUV 测量的误差范围内。