SPECT/CT商用设备中Tc-99m绝对定量的体模验证

Phantom Validation of Tc-99m Absolute Quantification in a SPECT/CT Commercial Device.

作者信息

Gnesin Silvano, Leite Ferreira Paulo, Malterre Jerome, Laub Priscille, Prior John O, Verdun Francis R

机构信息

Institute of Radiation Physics, Lausanne University Hospital, Lausanne, Switzerland.

Department of Nuclear Medicine and Molecular Imaging, Lausanne University Hospital, Lausanne, Switzerland.

出版信息

Comput Math Methods Med. 2016;2016:4360371. doi: 10.1155/2016/4360371. Epub 2016 Dec 14.

Abstract

. Similar to PET, absolute quantitative imaging is becoming available in commercial SPECT/CT devices. This study's goal was to assess quantitative accuracy of activity recovery as a function of image reconstruction parameters and count statistics in a variety of phantoms. . We performed quantitative Tc-SPECT/CT acquisitions (Siemens Symbia Intevo, Erlangen, Germany) of a uniform cylindrical, NEMA/IEC, and an anthropomorphic abdominal phantom. Background activity concentrations tested ranged: 2-80 kBq/mL. SPECT acquisitions used 120 projections (20 s/projection). Reconstructions were performed with the proprietary iterative conjugate gradient algorithm. NEMA phantom reconstructions were obtained as a function of the iteration number (range: 4-48). Recovery coefficients, hot contrast, relative lung error (NEMA phantom), and image noise were assessed. . In all cases, absolute activity and activity concentration were measured within 10% of the expected value. Recovery coefficients and hot contrast in hot inserts did not vary appreciably with count statistics. RC converged at 16 iterations for insert size > 22 mm. Relative lung errors were comparable to PET levels indicating the efficient integration of attenuation and scatter corrections with adequate detector modeling. . The tested device provided accurate activity recovery within 10% of correct values; these performances are comparable to current generation PET/CT systems.

摘要

与正电子发射断层扫描(PET)类似,绝对定量成像在商用单光子发射计算机断层扫描/计算机断层扫描(SPECT/CT)设备中也已可用。本研究的目的是评估在各种体模中,作为图像重建参数和计数统计函数的活度恢复的定量准确性。我们使用西门子Symbia Intevo(德国埃尔朗根)对均匀圆柱形、NEMA/IEC和拟人化腹部体模进行了定量锝(Tc)-SPECT/CT采集。测试的背景活度浓度范围为:2 - 80 kBq/mL。SPECT采集使用120个投影(每投影20秒)。使用专有的迭代共轭梯度算法进行重建。根据迭代次数(范围:4 - 48)获得NEMA体模的重建结果。评估了恢复系数、热对比度、相对肺误差(NEMA体模)和图像噪声。在所有情况下,绝对活度和活度浓度的测量值在预期值的10%以内。热插入物中的恢复系数和热对比度随计数统计没有明显变化。对于插入物尺寸> 22 mm,恢复系数在16次迭代时收敛。相对肺误差与PET水平相当,表明在适当的探测器建模下,衰减和散射校正得到了有效整合。测试的设备提供了准确的活度恢复,误差在正确值的10%以内;这些性能与当前一代PET/CT系统相当。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/f7f6/5206851/d01e7c72a6f0/CMMM2016-4360371.001.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验