Berman S, Hashizume M, Yang Y, DuPree J, Matsumoto T
Department of Surgery, Hahnemann University School of Medicine, Philadelphia, Pennsylvania 19102-1192.
Am J Surg. 1988 Mar;155(3):520-5. doi: 10.1016/s0002-9610(88)80127-2.
Intraoperative hemostasis and wound healing of 24 side-to-side intestinal anastomoses constructed with the ILA stapling device were studied in 12 dogs by comparing the ILA-32 and ILA-52 staple cartridges. Hemostasis was evaluated by intraoperative measurement of blood loss and bleeding time at the staple line. There was no statistically significant difference in mean blood loss (p greater than 0.05) or mean bleeding time (p greater than 0.10) between the two cartridges. Wound healing was studied using bursting strength measurements and silicone rubber casting of the microvasculature at the staple line. At 3 days, 1 week, and 2 weeks postoperatively, there was no significant difference between bursting strength values achieved with the two cartridges. Microscopic examination revealed that wound healing in the ILA-52 anastomoses lagged behind healing in the ILA-32 anastomoses at each postoperative time period studied. The silicone rubber casting study showed a paucity of microvasculature at the healing staple line with the ILA-52 cartridge as compared with the ILA-32 cartridge. Our findings suggest that the ILA-52 cartridge does not offer significantly improved intraoperative hemostasis over the ILA-32 cartridge and may affect the microvasculature at the staple line in a way that delays wound healing.
通过比较ILA-32和ILA-52钉仓,对12只犬构建的24个端端肠吻合术中的术中止血和伤口愈合情况进行了研究。通过术中测量钉合线处的失血量和出血时间来评估止血情况。两种钉仓之间的平均失血量(p大于0.05)或平均出血时间(p大于0.10)没有统计学上的显著差异。使用破裂强度测量和钉合线处微血管的硅橡胶铸型研究伤口愈合情况。术后3天、1周和2周,两种钉仓获得的破裂强度值没有显著差异。显微镜检查显示,在研究的每个术后时间段,ILA-52吻合口的伤口愈合都落后于ILA-32吻合口。硅橡胶铸型研究表明,与ILA-32钉仓相比,ILA-52钉仓在愈合的钉合线处微血管较少。我们的研究结果表明,ILA-52钉仓在术中止血方面并不比ILA-32钉仓有显著改善,并且可能以延迟伤口愈合的方式影响钉合线处的微血管。