Rosati R, Rebuffat C, Pezzuoli G
First Surgical Clinic, University of Milan, Italy.
Ann Surg. 1988 Mar;207(3):245-52. doi: 10.1097/00000658-198803000-00004.
The authors report the preliminary results obtained in animal and clinical experimentation of a new mechanical device for circular anastomosis which they have developed. It is a gun that places an apparatus consisting of three polypropylene rings that, through the compression among them of the severed edges of the bowel, realize a sutureless anastomosis and are spontaneously evacuated. Fifty-eight colonic anastomoses were performed in dogs with this device; 23 stapled colonic anastomoses were also executed concurrently. Forty-four animals underwent a relaparotomy to remove the colonic specimen containing the anastomoses. Bursting pressure and the histologic features of the anastomoses were evaluated at different time intervals after operation. A good healing of all compression anastomoses was observed, thereby allowing them to initiate the experience in humans. Thirteen anastomoses (6 colorectal extraperitoneal, 1 colorectal intraperitoneal, 5 colocolonic, 1 ileorectal) were performed at the 1st Surgical Department, Milan University. One subclinical leakage (7.7%) spontaneously healed in a few days. No stenoses were observed.
作者报告了他们研发的一种新型圆形吻合机械装置在动物和临床实验中获得的初步结果。它是一种枪式器械,可放置一个由三个聚丙烯环组成的装置,通过这三个环对肠管断端的挤压,实现无吻合钉吻合,且这些环会自行排出体外。使用该装置对58只犬进行了结肠吻合术;同时还进行了23例吻合钉结肠吻合术。44只动物接受了再次剖腹手术以切除包含吻合口的结肠标本。在术后不同时间间隔评估吻合口的破裂压力和组织学特征。观察到所有挤压吻合口愈合良好,从而得以在人体开展相关试验。米兰大学第一外科进行了13例吻合术(6例结直肠腹膜外吻合、1例结直肠腹膜内吻合、5例结肠结肠吻合、1例回直肠吻合)。1例亚临床漏(7.7%)在数天内自行愈合。未观察到狭窄。