运动单位分析。同心电极与单极电极的比较。

Motor unit analysis. Comparison between concentric and monopolar electrodes.

作者信息

Pease W S, Bowyer B L

机构信息

Department of Physical Medicine, Ohio State University, Columbus 43210.

出版信息

Am J Phys Med Rehabil. 1988 Feb;67(1):2-6.

PMID:3345237
Abstract

Motor unit action potential (MUAP) measurement parameters are reported to be affected by the type of needle electrode. MUAP amplitude measured with monopolar electrodes has been reported to be up to twice the amplitude recorded with concentric electrodes. We hypothesized that with current technology the difference in recordings from the two needles should be minimal and of little clinical significance. The right extensor digitorum communis muscle was examined in 15 healthy volunteers with first a concentric and then a monopolar needle. The filter band-pass of the electromyograph was 20 Hz to 10 kHz. Four first-recruited MUAPs were isolated by using each electrode. A separate muscle insertion was performed to locate each potential. Amplitudes were maximized. Consecutive MUAPs were averaged and measured. The mean (+/- SD) values of the measurements were as follows: amplitude: monopolar, 1038 (+/- 369) microV; concentric, 912 (+/- 315) microV; duration: monopolar, 10.8 (+/- 2.0) ms; concentric, 9.7 (+/- 1.7) ms. No statistically or clinically significant differences exist between these measurement values.

摘要

据报道,运动单位动作电位(MUAP)测量参数会受到针电极类型的影响。据报道,用单极电极测量的MUAP波幅高达同心电极记录波幅的两倍。我们推测,就目前的技术而言,两种针电极记录结果的差异应该很小,且临床意义不大。对15名健康志愿者的右侧指总伸肌进行了检查,先用同心针电极,然后用单极针电极。肌电图仪的滤波带通为20Hz至10kHz。使用每个电极分离出四个首次募集的MUAP。通过单独的肌肉进针来定位每个电位。将波幅最大化。对连续的MUAP进行平均并测量。测量的平均值(±标准差)如下:波幅:单极,1038(±369)μV;同心,912(±315)μV;时限:单极,10.8(±2.0)ms;同心,9.7(±1.7)ms。这些测量值之间不存在统计学或临床意义上的显著差异。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验