Andersen Andreas Gravgaard, Park Yang-Kyun, Elstrøm Ulrik Vindelev, Petersen Jørgen Breede Baltzer, Sharp Gregory C, Winey Brian, Dong Lei, Muren Ludvig Paul
Danish Centre for Particle Therapy, Aarhus University Hospital/Aarhus University, Aarhus, Denmark.
UT Southwestern Medical Center, Dallas, TX, USA.
Phys Imaging Radiat Oncol. 2020 Oct 27;16:89-94. doi: 10.1016/j.phro.2020.09.014. eCollection 2020 Oct.
Scatter correction of cone-beam computed tomography (CBCT) projections may enable accurate online dose-delivery estimations in photon and proton-based radiotherapy. This study aimed to evaluate the impact of scatter correction in CBCT-based proton range/dose calculations, in scans acquired in both proton and photon gantries.
CBCT projections of a Catphan and an Alderson phantom were acquired on both a proton and a photon gantry. The scatter corrected CBCTs (corrCBCTs) and the clinical reconstructions (stdCBCTs) were compared against CTs rigidly registered to the CBCTs (rigidCTs). The CBCTs of the Catphan phantom were segmented by materials for CT number analysis. Water equivalent path length (WEPL) maps were calculated through the Alderson phantom while proton plans optimized on the rigidCT and recalculated on all CBCTs were compared in a gamma analysis.
In medium and high-density materials, the corrCBCT CT numbers were much closer to those of the rigidCT than the stdCBCTs. E.g. in the 50% bone segmentations the differences were reduced from above 300 HU (with stdCBCT) to around 60-70 HU (with corrCBCT). Differences in WEPL from the rigidCT were typically well below 5 mm for the corrCBCTs, compared to well above 10 mm for the stdCBCTs with the largest deviations in the head and thorax regions. Gamma pass rates (2%/2mm) when comparing CBCT-based dose re-calculations to rigidCT calculations were improved from around 80% (with stdCBCT) to mostly above 90% (with corrCBCT).
Scatter correction leads to substantial artefact reductions, improving accuracy of CBCT-based proton range/dose calculations.
锥束计算机断层扫描(CBCT)投影的散射校正可实现基于光子和质子的放射治疗中准确的在线剂量输送估计。本研究旨在评估在质子和光子机架中采集的扫描中,散射校正在基于CBCT的质子射程/剂量计算中的影响。
在质子和光子机架上采集Catphan模体和Alderson模体的CBCT投影。将散射校正后的CBCT(corrCBCT)和临床重建图像(stdCBCT)与与CBCT进行刚性配准的CT(rigidCT)进行比较。通过材料对Catphan模体的CBCT进行分割以进行CT值分析。在Alderson模体中计算水等效路径长度(WEPL)图,同时比较在rigidCT上优化并在所有CBCT上重新计算的质子计划的伽马分析结果。
在中高密度材料中,corrCBCT的CT值比stdCBCT更接近rigidCT的CT值。例如,在50%骨分割中,差异从高于300 HU(使用stdCBCT)降低到约60 - 70 HU(使用corrCBCT)。与rigidCT相比,corrCBCT的WEPL差异通常远低于5毫米,而stdCBCT的差异远高于10毫米,在头部和胸部区域偏差最大。将基于CBCT的剂量重新计算与rigidCT计算进行比较时,伽马通过率(2%/2mm)从约80%(使用stdCBCT)提高到大多高于90%(使用corrCBCT)。
散射校正可大幅减少伪影,提高基于CBCT的质子射程/剂量计算的准确性。