Chełmiński Krzysztof, Bulski Wojciech
Medical Physics Department, The Maria Skłodowska-Curie Memorial Cancer Centre and Institute of Oncology, Roentgen Str. 5, 02781 Warsaw, Poland.
Phys Imaging Radiat Oncol. 2018 Apr 24;6:1-4. doi: 10.1016/j.phro.2018.03.001. eCollection 2018 Apr.
An audit methodology was developed and applied for output factor (OF) calculations in radiotherapy. The auditees were asked to calculate OFs for field sizes from 10 × 10 cm to 2 × 2 cm. Sixty five beams were audited; missing reference OFs were interpolated. The calculated OFs were in 73% of cases higher than the reference data. The smaller the field size, the higher the overestimations which were observed in the higher fraction of cases. Treatment planning systems generally overestimated OFs for small fields. The reference dataset helped radiotherapy centres to identify discrepancies which were higher than typical.
开发了一种审核方法并将其应用于放射治疗中的输出因子(OF)计算。要求被审核者计算10×10 cm至2×2 cm射野尺寸的OF。共审核了65束射野;缺失的参考OF通过插值法获得。在73%的病例中,计算得到的OF高于参考数据。射野尺寸越小,在更高比例的病例中观察到的高估情况越严重。治疗计划系统通常高估了小射野的OF。参考数据集有助于放射治疗中心识别高于典型情况的差异。