Department of Radiation Oncology, Brigham and Women's Hospital, Dana-Farber Cancer Institute, and Harvard Medical School, Boston, MA, 02115, United States of America.
Varian Medical Systems, Baden-Dattwil, CH-5405, Switzerland.
Phys Med Biol. 2021 Jun 21;66(13). doi: 10.1088/1361-6560/abddd2.
Simultaneous acquisition of cone beam CT (CBCT) projections using both the kV and MV imagers of an image guided radiotherapy system reduces set-up scan times-a benefit to lung cancer radiation oncology patients-but increases noise in the 3D reconstruction. In this article, we present a kV-MV scan time reduction technique that uses two noise-reducing measures to achieve superior performance. The first is a high-DQE multi-layer MV imager prototype. The second is a beam hardening correction algorithm which combines poly-energetic modeling with edge-preserving, regularized smoothing of the projections. Performance was tested in real acquisitions of the Catphan 604 and a thorax phantom. Percent noise was quantified from voxel values in a soft tissue volume of interest (VOI) while edge blur was quantified from a VOI straddling a boundary between air and soft material. Comparisons in noise/resolution performance trade-off were made between our proposed approach, a dose-equivalent kV-only scan, and a kV-MV reconstruction technique previously published by Yin(20059). The proposed technique demonstrated lower noise as a function of spatial resolution than the baseline kV-MV method, notably a 50% noise reduction at typical edge blur levels. Our proposed method also exhibited fainter non-uniformity artifacts and in some cases superior contrast. Overall, we find that the combination of a multi-layer MV imager, acquiring at a LINAC source energy of 2.5 MV, and a denoised beam hardening correction algorithm enables noise, resolution, and dose performance comparable to standard kV-imager only set-up CBCT, but with nearly half the gantry rotation time.
同时使用图像引导放射治疗系统的千伏 (kV) 和兆伏 (MV) 成像仪获取锥形束 CT (CBCT) 投影可减少设置扫描时间——这对肺癌放射肿瘤学患者有益——但会增加 3D 重建中的噪声。在本文中,我们提出了一种 kV-MV 扫描时间缩短技术,该技术使用两种降噪措施来实现卓越的性能。第一个是高 DQE 多层 MV 成像仪原型。第二个是束硬化校正算法,它结合了多能建模和边缘保持、投影正则化平滑。性能在 Catphan 604 和胸部体模的实际采集过程中进行了测试。在软组织感兴趣体积 (VOI) 中的体素值中量化了噪声百分比,而在跨越空气和软组织边界的 VOI 中量化了边缘模糊度。对我们提出的方法、剂量等效 kV 单扫描和 Yin(20059)之前发表的 kV-MV 重建技术之间的噪声/分辨率性能权衡进行了比较。与基线 kV-MV 方法相比,所提出的技术表现出更低的噪声作为空间分辨率的函数,特别是在典型的边缘模糊水平下噪声降低了 50%。我们提出的方法还表现出更微弱的非均匀伪影,在某些情况下对比度更高。总体而言,我们发现多层 MV 成像仪的组合,以 LINAC 源能量 2.5 MV 进行采集,以及去噪的束硬化校正算法可实现与标准 kV 成像仪仅设置 CBCT 相当的噪声、分辨率和剂量性能,但旋转时间缩短了近一半。