Department of Radiation Oncology, University of Colorado School of Medicine, Aurora, CO 80045, United States of America.
Phys Med Biol. 2022 Aug 9;67(16). doi: 10.1088/1361-6560/ac8268.
. The concept of using kilovoltage (kV) and megavoltage (MV) beams concurrently has potential applications in cone beam computed tomography (CBCT) guided radiation therapy, such as single breath hold scans, metal artifact reduction, and simultaneous imaging during MV treatment delivery. However, MV cross-scatter generated during MV beam delivery degrades CBCT image quality. To address this, a 2D antiscatter grid and a cross-scatter correction method were investigated in the context of high dose MV treatment delivery.. A 3D printed, tungsten 2D antiscatter grid prototype was utilized in kV CBCT scans to reduce MV cross-scatter fluence during concurrent MV beam delivery. Remaining cross-scatter in projections was corrected by using the 2D grid itself as a cross-scatter intensity sampling device, referred to as grid-based scatter sampling (GSS). To test this approach, kV CBCT acquisitions were performed while delivering 6 and 10 MV beams, mimicking high dose rate treatment delivery scenarios. kV and MV beam deliveries were not synchronized to eliminate MV beam delivery interruption. MV cross-scatter suppression performance of the proposed approach was evaluated in projections and CBCT images of phantoms.. 2D grid reduced the intensity of MV cross-scatter in kV projections by a factor of 3 on the average, when compared to conventional antiscatter grid. Remaining cross scatter as measured by the GSS method was within 7% of measured reference intensity values, and subsequently corrected. CBCT image quality was improved substantially during concurrent kV-MV beam delivery. Median Hounsfield Unit (HU) inaccuracy was up to 182 HU without our methods, and it was reduced to a median 6.5 HU with our 2D grid and scatter correction approach. Our methods provided a factor of 2-6 improvement in contrast-to-noise ratio.. This investigation demonstrates the utility of 2D antiscatter grids and grid-based scatter sampling in suppressing MV cross-scatter. Our approach successfully minimized the effects of MV cross-scatter in concurrent kV CBCT imaging and high dose MV treatment delivery scenarios. Hence, robust MV cross-scatter suppression is potentially feasible without MV beam delivery interruption or compromising kV image acquisition rate.
. 使用千伏 (kV) 和兆伏 (MV) 射线束同时进行的概念在锥形束计算机断层扫描 (CBCT) 引导的放射治疗中有潜在的应用,例如单次屏气扫描、减少金属伪影以及在 MV 治疗递送期间进行同时成像。然而,MV 射线束递送期间产生的 MV 散射会降低 CBCT 图像质量。为了解决这个问题,在高剂量 MV 治疗递送的背景下研究了二维散射栅和散射校正方法。使用 3D 打印的钨二维散射栅原型在 kV CBCT 扫描中减少 MV 射线束同时递送期间的 MV 散射强度。通过使用二维栅格本身作为散射强度采样装置,即基于栅格的散射采样 (GSS),对投影中的剩余散射进行校正。为了测试这种方法,在模拟高剂量率治疗递送场景的情况下进行了 6 和 10 MV 射线束的 kV CBCT 采集。kV 和 MV 射线束的递送没有同步,以消除 MV 射线束递送中断。在体模的 CBCT 图像和投影中评估了所提出方法的 MV 散射抑制性能。. 与传统的散射栅相比,二维栅格将 kV 投影中的 MV 散射强度平均降低了 3 倍。通过 GSS 方法测量的剩余散射在测量参考强度值的 7%以内,随后进行了校正。在同时进行 kV-MV 射线束递送期间,CBCT 图像质量得到了显著改善。没有我们的方法时,中位数 Hounsfield 单位 (HU) 不准确高达 182 HU,而使用我们的二维栅格和散射校正方法后,中位数 HU 不准确降低到 6.5 HU。我们的方法提供了对比度噪声比的 2-6 倍的改善。. 这项研究证明了二维散射栅和基于栅格的散射采样在抑制 MV 散射方面的实用性。我们的方法成功地最小化了同时进行的 kV CBCT 成像和高剂量 MV 治疗递送场景中 MV 散射的影响。因此,在不中断 MV 射线束递送或不影响 kV 图像采集速率的情况下,稳健的 MV 散射抑制是有可能实现的。