Prospective dual-surrogate validation study of periodic imaging during treatment for accurately monitoring intrafraction motion of prostate cancer patients.

机构信息

The Ohio State University, Department of Radiation Oncology, Columbus, United States.

The Ohio State University, Department of Radiation Oncology, Columbus, United States.

出版信息

Radiother Oncol. 2021 Apr;157:40-46. doi: 10.1016/j.radonc.2021.01.020. Epub 2021 Jan 21.

Abstract

BACKGROUND AND PURPOSE

The goal of this prospective study is to validate the use of periodic imaging during treatment with a fiducial marker detection algorithm using radiofrequency transponders for prostate cancer patients undergoing treatment for radiation therapy.

MATERIALS AND METHODS

Ten male patients were enrolled in this study and treated for prostate cancer with implanted electromagnetic monitoring beacons. We evaluated the accuracy and limitations of Intrafraction Motion Review (IMR) by comparing the known locations of the beacons using the electromagnetic monitoring system to the position data reported from IMR images.

RESULTS

A total of 4054 images were taken during treatment. The difference in vector magnitude of the two methods is centered around zero (mean: 0.03 cm, SD: 0.16 cm) and Lin's Concordance Correlation Coefficient (CCC) is 0.99 (95% CI: 0.98, 1) overall. The Euclidean distance between the two methods was close to zero (median: 0.09 cm, IQR: 0.06, 0.14 cm). The difference in distance between any two markers was centered around zero (mean: 0.01 cm, SD: 0.12 cm) and Lin's CCC is 0.97 (95% CI: 0.96, 0.98) overall.

CONCLUSION

The accuracy of the algorithm for detected markers within the 2D images is comparable to electromagnetic monitoring for fiducial identification when detected. IMR could provide an alternate solution for patients with contraindications of use of an electromagnetic monitoring system and a cost effective alternative to the acquisition of an additional system for patient monitoring, but does not provide data for pre-treatment set-up verification and real-time 3D positioning during treatment.

摘要

背景与目的

本前瞻性研究的目的是验证在接受放射治疗的前列腺癌患者中,使用带有射频应答器的基准标记物检测算法进行治疗期间周期性成像的实用性。

材料与方法

本研究纳入了 10 名男性患者,他们因前列腺癌接受了植入式电磁监测信标治疗。我们通过将电磁监测系统中已知信标位置与来自 IMR 图像的位置数据进行比较,评估了 Intrafraction Motion Review(IMR)的准确性和局限性。

结果

在治疗过程中共拍摄了 4054 张图像。两种方法的向量幅度差异集中在零附近(均值:0.03cm,标准差:0.16cm),Lin 的一致性相关系数(CCC)总体为 0.99(95%置信区间:0.98,1)。两种方法之间的欧几里得距离接近零(中位数:0.09cm,IQR:0.06,0.14cm)。任何两个标记物之间的距离差异集中在零附近(均值:0.01cm,标准差:0.12cm),Lin 的 CCC 总体为 0.97(95%置信区间:0.96,0.98)。

结论

当检测到标记物时,算法在 2D 图像中检测到的标记物的准确性可与电磁监测相媲美,用于基准识别。IMR 可为电磁监测系统使用禁忌的患者提供替代解决方案,并且是一种具有成本效益的替代方案,无需为患者监测获取额外系统,但它不能提供治疗前设置验证和治疗期间实时 3D 定位的数据。