IRCCS Istituto Auxologico Italiano, Obesity Unit and Laboratory of Nutrition and Obesity Research, Department of Endocrine and Metabolic Diseases, Milan, Italy; International Center for the Assessment of Nutritional Status, Department of Food, Environmental and Nutritional Sciences, University of Milan, Milan, Italy.
International Center for the Assessment of Nutritional Status, Department of Food, Environmental and Nutritional Sciences, University of Milan, Milan, Italy.
Clin Nutr ESPEN. 2021 Feb;41:443-446. doi: 10.1016/j.clnesp.2020.10.015. Epub 2020 Nov 10.
BACKGROUND & AIMS: During the coronavirus disease 2019 (COVID-19) pandemic the use of Indirect calorimetry (IC) during nutritional rehabilitation program requires special precautions due to possible contagions for patients and health professionals. We evaluated in a sample of healthy subjects the agreement between oxygen consumption (VO mL/min), carbon dioxide production (VCO mL/min), respiratory quotient (RQ) and resting energy expenditure (REE kcal/24 h/day) measured by IC with and without a filtering facepiece mask.
10 subjects with a mean (SD) age of 43 (10) years and a body mass index of 25.2 (5.8) kg/m underwent indirect calorimetry both with and without a class 2 filtering facepiece mask (FFP2), in random order. The limits of agreement (LOA) and the concordance correlation coefficient (CCC) were used to evaluate the interchangeability of the measurement conditions.
The LOA between REE measured with and without FFP2 (-111 to 189 kcal/day) were comparable to those for repeated IC tests without wearing masks and CCC (0.95) showed substantial agreement.
We observed high agreement between REE measured by IC with and without FFP2 mask. These procedures are interchangeable in clinical practice.