Department of Radiology, Zuyderland Medical Center, Heerlen/Sittard/Geleen, The Netherlands.
Neuroradiology. 2021 Oct;63(10):1611-1616. doi: 10.1007/s00234-021-02659-1. Epub 2021 Feb 3.
To evaluate the reliability and accuracy of thick maximum intensity projection (MIP) CTA images to detect large-vessel occlusion (LVO) in the anterior circulation in patients with acute stroke.
A total of 140 acute stroke patients (41 with and 99 without LVO) were evaluated by two neuroradiologists for LVO using axial 3-mm and 2-mm MIPs.
Interobserver agreement was substantial using 3-mm MIPs (ĸ = 0.67) and almost perfect using 2-mm MIPs (ĸ = 0.82). Using 3-mm MIPs, sensitivities were 80.5% and 68.3%, with specificities of 98.0% and 96.0%. Using 2-mm MIPs, sensitivities were 82.9% and 73.2%, with specificities of 98.0% and 99.0%. Sensitivity and specificity of 3 mm and 2 mm MIPs were not statistically significantly different (P ≥ 0.375). The majority of LVOs in the distal intracranial carotid artery, and/or M1-segment were correctly identified: 96.0% (observer 1, 3-mm MIPs), 88.0% (observer 2, 3-mm MIPs), 96.0% (observer 1, 2-mm MIPs), and 96.0% (observer 2, 2 mm MIPs). Using 3-mm MIP images, observers 1 and 2 missed 7/15 (46.7%) and 9/15 (60.0%) of isolated M2-segment occlusions, respectively. Using 2-mm MIP images, observers 1 and 2 missed 5/15 (33.3%) and 6/15 (40.0%) of isolated M2-segment occlusions, respectively.
Thick (2-3 mm) axial MIPs are not useful to detect proximal LVO in the anterior circulation.
评估厚最大密度投影(MIP)CTA 图像在检测急性脑卒中前循环大血管闭塞(LVO)中的可靠性和准确性。
共对 140 例急性脑卒中患者(41 例有 LVO 和 99 例无 LVO)进行了轴向 3 毫米和 2 毫米 MIP 的两位神经放射科医生 LVO 评估。
使用 3 毫米 MIP 时,观察者间的一致性为中等(ĸ=0.67),使用 2 毫米 MIP 时,观察者间的一致性几乎完美(ĸ=0.82)。使用 3 毫米 MIP,灵敏度分别为 80.5%和 68.3%,特异性分别为 98.0%和 96.0%。使用 2 毫米 MIP,灵敏度分别为 82.9%和 73.2%,特异性分别为 98.0%和 99.0%。3 毫米和 2 毫米 MIP 的敏感性和特异性无统计学差异(P≥0.375)。远端颈内动脉和/或 M1 段的大多数 LVO 均被正确识别:96.0%(观察者 1,3 毫米 MIPs),88.0%(观察者 2,3 毫米 MIPs),96.0%(观察者 1,2 毫米 MIPs),和 96.0%(观察者 2,2 毫米 MIPs)。使用 3 毫米 MIP 图像,观察者 1 和 2 分别漏诊了 7/15(46.7%)和 9/15(60.0%)孤立的 M2 段闭塞。使用 2 毫米 MIP 图像,观察者 1 和 2 分别漏诊了 5/15(33.3%)和 6/15(40.0%)孤立的 M2 段闭塞。
厚(2-3 毫米)轴位 MIP 对于检测前循环近端 LVO 没有帮助。