Preventing nickel-titanium rotary instrument from breakage by continuous irrigation with different fluids during root canal preparation.

作者信息

Li Changjian, Liu Le, Li Yanfeng, Wang Xin, Liu Haiyun, Bao Jiaxin, Wang Jing, Fan Jiadong

机构信息

Department of Stomatology, the Fourth Medical Center, Chinese PLA General Hospital, Beijing, China.

出版信息

Medicine (Baltimore). 2021 Jan 22;100(3):e23865. doi: 10.1097/MD.0000000000023865.

Abstract

To study the effect of continuous irrigation of rotating nickel-titanium instrument with several common clinical fluids on the diameter, breaking length and breaking position of nickel-titanium instrument, so as to provide some reference and theoretical basis for clinical operation and instrument improvement.A standardized curved root canal model was established, and ProTaper Universal (PTU) F1 instrument was selected for root canal preparation. The nickel-titanium F1 instrument was flushed with distilled water, 0.9% NaCl, 0.2% chlorhexidine, 1% sodium hypochlorite and 5% sodium hypochlorite, and the diameter, length and position of the instrument before and after breakage were recorded.Only 5% sodium hypochlorite influenced the diameter of 6 mm marker points under different irrigation conditions (P < .05). There was no statistical difference in the length of broken instruments among all the groups, and torsional deformation mainly occurred at the end of broken instruments. The broken positions of instruments in all the groups were located at the bending segment of the root canal. The breaking frequency of the 5% sodium hypochlorite group was the highest in the area 3-5.5 mm away from apical foramen, while the other 4 groups had the highest breaking frequency in the area 0 to 1.5 mm away from apical foramen.External irrigation with different fluids did not influence the breaking length of instruments. The closer to the apical foramen was, the higher the breaking frequency of instruments was. However, only 5% sodium hypochlorite can affect the diameter of rotary nickel-titanium instruments, and may lead to early breakage of the instrument, indicating that the use of disinfectants, except 5% sodium hypochlorite, cannot reduce breakage resistance of nickel-titanium instrument compared with distilled water flushing. Furthermore, 5% hypochlorite could not be recommended for irrigation in clinical practice.

摘要

研究几种临床常用冲洗液持续冲洗旋转镍钛器械对镍钛器械直径、折断长度及折断位置的影响,为临床操作及器械改进提供参考和理论依据。建立标准化弯曲根管模型,选用ProTaper Universal(PTU)F1器械进行根管预备。将镍钛F1器械分别用蒸馏水、0.9%氯化钠、0.2%氯己定、1%次氯酸钠和5%次氯酸钠冲洗,记录器械折断前后的直径、长度及位置。仅5%次氯酸钠在不同冲洗条件下影响6mm标记点处的直径(P<0.05)。所有组折断器械的长度无统计学差异,扭转变形主要发生在折断器械的末端。所有组器械的折断位置均位于根管弯曲段。5%次氯酸钠组在距根尖孔3~5.5mm区域折断频率最高,而其他4组在距根尖孔0~1.5mm区域折断频率最高。不同冲洗液进行根管冲洗对器械折断长度无影响。器械距根尖孔越近,折断频率越高。然而,仅5%次氯酸钠可影响旋转镍钛器械的直径,并可能导致器械早期折断,表明除5%次氯酸钠外,使用消毒剂与蒸馏水冲洗相比不能降低镍钛器械的抗折断性能。此外,5%次氯酸钠不推荐在临床操作中用于冲洗。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/1e4c/7837962/94834af244fa/medi-100-e23865-g001.jpg