Department of Radiology, Kochi Medical School Hospital, Kochi, Japan.
Departments of Gunma Prefectural College of Health Sciences, Maebashi, Japan.
Nucl Med Rev Cent East Eur. 2021;24(1):1-10. doi: 10.5603/NMR.2021.0001.
The authors aimed to elucidate the effect of liver space-occupying lesions (SOL) on the quantitative index of the hepatic reserve, calculated using the dynamic planar image (LHLplanar), and a three-dimensional quantitative index (LHLSPECT) calculated using quantitative combined modality single-photon emission computed tomography (SPECT/CT).
Water balloons of different volumes that simulated liver SOL were placed in various positions in the combined cardiac-liver phantom to examine the effects of liver SOL on visualization and quantitative indicators (LHLplanar and LHLSPECT). A 200 mL water balloon was placed in the anterior right, posterior right, left medial and left lateral lobes in the liver phantom to compare LHLplanar and LHLSPECT values with and without liver SOL at each position. Subsequently, volumes of those in the front of the right lobe were changed to 50 mL, 100 mL, 200 mL, and 400 mL, followed by statistically comparing LHLplanar and LHLSPECT values in the presence and absence of liver SOL.
Despite the variation in the degree of defect accumulation with the location of the balloon when using frontal planar imaging, quantitative SPECT/CT imaging identified all defects. Multiple comparison analysis revealed that unlike LHLSPECT, the LHLplanar values changed according to liver SOL position and volume.
Liver SOL position and volume may affect the hepatic reserve assessments performed using LHLplanar values. In contrast, LHLSPECT is calculated using quantitative SPECT/CT and considers the effects of scattering and attenuation corrections. Therefore, LHLSPECT is a more accurate quantitative indicator of hepatic reserve than LHLplanar and is expected to facilitate future clinical research.
本研究旨在阐明肝脏占位性病变(SOL)对使用动态平面图像(LHLplanar)计算的肝脏储备定量指标和使用定量组合模态单光子发射计算机断层扫描(SPECT/CT)计算的三维定量指标(LHLSPECT)的影响。
在心脏-肝脏组合体模中,将不同体积的水球放置在不同位置以模拟 SOL,以研究 SOL 对可视化和定量指标(LHLplanar 和 LHLSPECT)的影响。在体模的右前、右后、左内和左外侧叶分别放置一个 200ml 的水球,比较每个位置有无 SOL 时的 LHLplanar 和 LHLSPECT 值。随后,改变右前叶水球的体积为 50ml、100ml、200ml 和 400ml,并统计比较有无 SOL 时的 LHLplanar 和 LHLSPECT 值。
尽管使用前位平面成像时,随着气球位置的不同,缺损程度的积累存在差异,但定量 SPECT/CT 成像可识别所有缺陷。多重比较分析表明,与 LHLSPECT 不同,LHLplanar 值随 SOL 位置和体积的变化而变化。
SOL 的位置和体积可能会影响使用 LHLplanar 值进行的肝脏储备评估。相比之下,LHLSPECT 使用定量 SPECT/CT 计算,并考虑了散射和衰减校正的影响。因此,LHLSPECT 是一种比 LHLplanar 更准确的肝脏储备定量指标,有望促进未来的临床研究。