Brown S A, Barsanti J A
Department of Physiology and Pharmacology, College of Veterinary Medicine, University of Georgia, Athens 30602.
Am J Vet Res. 1988 Mar;49(3):321-4.
A quantitative buffy coat (QBC) analysis was evaluated for 175 canine, 125 feline, and 125 equine blood samples. The method used centrifuged whole blood and yielded rapid results expressed as respective band lengths for RBC, granulocytes, nongranulocytes, and platelets. Simple regression analysis of band lengths and reference laboratory methods yielded correlation coefficients (r) ranging from 0.72 to 0.99. The PCV, granulocyte count, and total WBC count, as determined by the 2 methods, correlated well (r greater than or equal to 0.93 in all cases). Platelet and nongranulocyte counts were less well correlated. The QBC system provided a means of performing rapid hematologic screening. The principal problem encountered was poor separation of the RBC-granulocyte interface in 17% of canine samples, which interfered with measurement of band lengths. Evaluation of the QBC tube for detection of Dirofilaria immitis microfilaremia revealed 100% sensitivity to counts as low as 160 microfilariae/ml of whole blood.
对175份犬类、125份猫类和125份马类血液样本进行了定量血沉棕黄层(QBC)分析。该方法采用全血离心,结果以红细胞、粒细胞、非粒细胞和血小板的各自条带长度快速呈现。条带长度与参考实验室方法的简单回归分析得出相关系数(r)在0.72至0.99之间。两种方法测定的红细胞压积、粒细胞计数和白细胞总数相关性良好(所有情况下r均大于或等于0.93)。血小板和非粒细胞计数的相关性较差。QBC系统提供了一种进行快速血液学筛查的方法。遇到的主要问题是17%的犬类样本中红细胞-粒细胞界面分离不佳,这干扰了条带长度的测量。对用于检测犬恶丝虫微丝蚴血症的QBC管的评估显示,对低至160条微丝蚴/毫升全血的计数灵敏度为100%。