机械肺振动器和高频胸壁振荡在植入式心脏电子设备患者中的安全性。

Safety of mechanical lung vibrator and high-frequency chest wall oscillation in patients with cardiac implantable electronic device.

机构信息

Division of Cardiology, Department of Internal Medicine, Samsung Changwon Hospital, Sungkyunkwan University School of Medicine, Changwon, Republic of Korea.

Division of Cardiology, Department of Internal Medicine, Heart Vascular Stroke Institute, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, Republic of Korea.

出版信息

Clin Cardiol. 2021 Apr;44(4):531-536. doi: 10.1002/clc.23571. Epub 2021 Feb 16.

Abstract

BACKGROUND

Chest physiotherapy (CPT) is a non-pharmacological therapy to facilitate airway secretion removal. There have been concerns about potential electromagnetic interference (EMI) and lead integrity problems during the use of vibrating CPT devices in patients with cardiac implantable electronic devices (CIEDs).

HYPOTHESIS

Two CPT devices can be used safely in patients with CIED.

METHODS

Volunteer patients with CIED underwent device interrogation to check lead integrity and device function before and after application of CPT devices. Mechanical lung vibrator and high-frequency chest wall oscillation (HFCWO) vests were used while monitoring surface electrocardiograms and intra-cardiac electrograms.

RESULTS

We prospectively enrolled 46 patients with CIEDs (25 pacemakers, 15 implantable cardioverter-defibrillators, and six cardiac resynchronization therapy-defibrillators). There was no noise detection or EMI during CPT in any patient. None of the patients showed clinically significant changes in lead integrity parameters. HFCWO inappropriately accelerated the pacing rate up to the maximal programmed value in five patients with pacemakers and two with cardiac resynchronization therapy-defibrillators.

CONCLUSION

CPT may be safely applied to patients with CIED without compromising lead integrity or device function, except for unwanted increase in pacing rate caused by misdetection of chest wall vibration as patients' activity while using HFCWO. Deactivation of the accelerometer-based activity sensor may be needed when HFCWO is planned for CPT.

摘要

背景

胸部物理疗法(CPT)是一种促进气道分泌物清除的非药物治疗方法。在使用振动式 CPT 设备治疗患有心脏植入式电子设备(CIED)的患者时,人们一直担心潜在的电磁干扰(EMI)和导联完整性问题。

假设

两种 CPT 设备可安全应用于患有 CIED 的患者。

方法

志愿者 CIED 患者在应用 CPT 设备前后进行设备检查,以检查导联完整性和设备功能。使用机械肺振动器和高频胸壁振荡(HFCWO)背心,同时监测体表心电图和心内电图。

结果

我们前瞻性纳入了 46 例 CIED 患者(25 例起搏器,15 例植入式心脏复律除颤器和 6 例心脏再同步治疗除颤器)。在任何患者中,CPT 期间均未检测到噪声或 EMI。没有患者表现出导联完整性参数的临床显著变化。HFCWO 不当加速了 5 例起搏器和 2 例心脏再同步治疗除颤器患者的起搏率,直至最大程控值。

结论

CPT 可安全应用于 CIED 患者,而不会损害导联完整性或设备功能,但在使用 HFCWO 时,由于误检测患者活动引起的胸壁振动,可能会导致起搏率意外增加。在计划使用 HFCWO 进行 CPT 时,可能需要停用基于加速度计的活动传感器。