Wang Marlene D, Truong Cecile, Mammo Zaid, Hussnain Syed Amal, Chen Royce W S
Department of Ophthalmology, Edward S. Harkness Eye Institute, Columbia University Irving Medical Center, New York, NY, 10032, USA.
Clin Ophthalmol. 2021 Feb 11;15:507-512. doi: 10.2147/OPTH.S297307. eCollection 2021.
Biomicroscopy, B-scan ultrasound imaging, and SD-OCT are all modalities used to characterize a posterior vitreous detachment (PVD). Our objective is to assess the precision of the diagnosis of PVD by SS-OCT.
This prospective observational study examines ninety-five eyes of forty-nine patients with biomicroscopy, B-scan ultrasound, and SS-OCT for the presence or absence of a complete PVD. All SS-OCT images were reviewed by two retina specialists (RWSC, ZM). All three diagnostic methods were evaluated for agreement by Cohen's kappa statistic.
The inter-rater reliability between retina specialists reading the SS-OCT images was 97.9% (κ = 0.957). Agreement on PVD status between SS-OCT and biomicroscopy was 85.3% (κ = 0.711). Agreement between SS-OCT and B-scan ultrasound was 83.2% (κ = 0.667). Agreement between B-scan ultrasound and biomicroscopy was 87.4% (κ = 0.743).
For the diagnosis of complete PVD, SS-OCT allows for high accuracy and agreement between graders.
生物显微镜检查、B 超超声成像和 SD-OCT 都是用于描述玻璃体后脱离(PVD)的方法。我们的目的是评估 SS-OCT 诊断 PVD 的准确性。
这项前瞻性观察性研究对 49 例患者的 95 只眼睛进行了生物显微镜检查、B 超超声检查和 SS-OCT 检查,以确定是否存在完全性 PVD。所有 SS-OCT 图像均由两名视网膜专家(RWSC、ZM)进行评估。采用 Cohen's kappa 统计量评估这三种诊断方法的一致性。
阅读 SS-OCT 图像的视网膜专家之间的评分者间可靠性为 97.9%(κ = 0.957)。SS-OCT 与生物显微镜检查在 PVD 状态上的一致性为 85.3%(κ = 0.711)。SS-OCT 与 B 超超声检查之间的一致性为 83.2%(κ = 0.667)。B 超超声检查与生物显微镜检查之间的一致性为 87.4%(κ = 0.743)。
对于完全性 PVD 的诊断,SS-OCT 具有较高的准确性,且分级者之间的一致性良好。