便携式压力测量装置与PicoPress®在压缩疗法期间对肥厚性烧伤瘢痕进行服装压力测量的比较。

Comparison between the portable pressure measuring device and PicoPress® for garment pressure measurement on hypertrophic burn scar during compression therapy.

作者信息

Lee Seung Yeol, Cho Yoon Soo, Joo So Young, Seo Cheong Hoon

机构信息

Department of Physical Medicine and Rehabilitation, Soonchunhyang University Bucheon Hospital, Soonchunhyang University College of Medicine, Bucheon, Republic of Korea.

Department of Rehabilitation Medicine, Hangang Sacred Heart Hospital, College of Medicine, Hallym University, Seoul, Republic of Korea.

出版信息

Burns. 2021 Nov;47(7):1621-1626. doi: 10.1016/j.burns.2021.01.018. Epub 2021 Feb 9.

Abstract

PURPOSE

The current standard treatment for hypertrophic scars following burn injury is pressure garment therapy. The experimenters developed the novel portable pressure measuring device using silicon piezoresistive sensors. As PicoPress® is the most accurate (i.e., lowest variation and error) manometric sensor for pressure measurement, we sought to compare and examine the accuracy of the novel device regarding in vitro pressure measurements at the hypertrophic scar-pressure garment interface.

METHODS

The novel device was designed to operate in non-corrosive media, such as air. The device can use up to six pressure sensing points and was developed to adjust the number of pressure sensors according to the size of the scar. Pressure measurements were acquired through a readout circuit consisting of an analog-to-digital converter, a microprocessor, and a Bluetooth transmission module for wireless data transmission to an external device. All signals were converted into mean pressure expressed in millimeters of mercury (mmHg). The mean pressure values measured by the sensors were compared to those obtained from PicoPress®. 55 garment pressures recordings were obtained from the sensors over this study conducted in 2018-February 2020. We then analyzed the test-retest reliability using the intraclass correlation coefficients (ICC). PicoPress® was also employed in the same pressure garments for obtaining similar measurements. A two way random effects model ICC with 95% confidence intervals was used to compare the mean pressure values obtained from the silicon piezoresistive sensors to the PicoPress® measurements.

RESULTS

The test-retest reliability of the pressure sensors was close to the acceptable level for clinical use regarding stationary interface pressure measurement (ICC = 0.99, 95% CI 0.990-0.997). The mean pressure obtained from the silicon piezoresistive pressure sensors showed an accordance with the measurements from PicoPress® (ICC = 0.97, 95% CI 0.947-0.985).

CONCLUSION

The novel device may present a viable alternative to PicoPress® for garment pressure measurements. In addition, the novel device improves adaptability to the hypertrophic scar shape and size. Complementary characteristics such as wireless transmission to an external device may allow burn patients to continuously wear the device for real-time measurements during pressure garment therapy, thus improving existing devices including PicoPress®.

摘要

目的

烧伤后增生性瘢痕的当前标准治疗方法是压力衣疗法。实验者开发了一种使用硅压阻式传感器的新型便携式压力测量装置。由于PicoPress®是用于压力测量的最精确(即变化和误差最小)的压力传感器,我们试图比较并检验该新型装置在增生性瘢痕 - 压力衣界面进行体外压力测量时的准确性。

方法

该新型装置设计用于在非腐蚀性介质(如空气)中运行。该装置最多可使用六个压力传感点,并根据瘢痕大小开发了调整压力传感器数量的功能。压力测量通过一个由模数转换器、微处理器和用于无线数据传输到外部设备的蓝牙传输模块组成的读出电路进行。所有信号均转换为以毫米汞柱(mmHg)表示的平均压力。将传感器测量的平均压力值与从PicoPress®获得的值进行比较。在2018年 - 2020年2月进行的这项研究中,从传感器获得了55次压力衣压力记录。然后我们使用组内相关系数(ICC)分析重测信度。PicoPress®也用于相同的压力衣以获得类似测量值。使用具有95%置信区间的双向随机效应模型ICC来比较从硅压阻式传感器获得的平均压力值与PicoPress®测量值。

结果

对于静态界面压力测量,压力传感器的重测信度接近临床使用的可接受水平(ICC = 0.99,95% CI 0.990 - 0.997)。从硅压阻式压力传感器获得的平均压力与PicoPress®的测量值一致(ICC = 0.97,95% CI 0.947 - 0.985)。

结论

该新型装置可能是用于压力衣压力测量的PicoPress®的可行替代方案。此外,该新型装置提高了对增生性瘢痕形状和大小的适应性。诸如无线传输到外部设备等补充特性可能使烧伤患者在压力衣治疗期间能够持续佩戴该装置进行实时测量,从而改进包括PicoPress®在内的现有装置。