Suppr超能文献

Pliance X在测量儿童压力衣与烧伤瘢痕界面压力方面的可靠性及临床实用性。

Reliability and clinical utility of the Pliance X for measuring pressure at the interface of pressure garments and burn scars in children.

作者信息

Wiseman Jodie, Simons Megan, Kimble Roy, Tyack Zephanie

机构信息

Centre for Children's Burns and Trauma Research, Children's Health Research Centre, The University of Queensland, Brisbane, QLD, Australia.

Pegg Leditschke Children's Burns Centre, Lady Cilento Children's Hospital, Brisbane, QLD, Australia.

出版信息

Burns. 2018 Nov;44(7):1820-1828. doi: 10.1016/j.burns.2018.05.002. Epub 2018 Aug 1.

Abstract

BACKGROUND

Reliable, clinically acceptable pressure measuring devices are important to accurately record interface pressure. The Pliance X is a device that may overcome previous device limitations (i.e. pressure garment distortion, temperature dependent sensors). This research aimed to identify the test-retest reliability and clinical utility of the Pliance X in children receiving pressure garment therapy post-burn.

METHODS

Data was collected as part of a pilot longitudinal cohort study. Immediate test-retest reliability was investigated at baseline and analysed using Intra-Class Correlation Coefficients (ICCs). Clinical utility was evaluated at baseline using the criteria of appropriate, accessible, practicable and acceptable.

RESULTS

Thirty-one children completed baseline measurements. Participants had a median age of three years (IQR: 1.5, 7.5), and 61% were male. Test-retest reliability of the Pliance X was close to acceptable for clinical use for stationary interface pressure (ICC=0.87, n=27). The Pliance X was appropriate and acceptable for children of all ages who were not distressed.

DISCUSSION

Further reliability testing is required, including inter-rater reliability testing. The device's clinical utility could be improved with alternate sensor styles and an instant mean pressure reading to enable immediately clinically useful information on the interface pressure.

摘要

背景

可靠且临床可接受的压力测量设备对于准确记录界面压力很重要。Pliance X是一种可能克服先前设备局限性(即压力衣变形、温度依赖传感器)的设备。本研究旨在确定Pliance X在烧伤后接受压力衣治疗的儿童中的重测信度和临床效用。

方法

数据作为一项纵向队列研究试点的一部分进行收集。在基线时调查即时重测信度,并使用组内相关系数(ICC)进行分析。在基线时使用适当、可及、可行和可接受的标准评估临床效用。

结果

31名儿童完成了基线测量。参与者的中位年龄为3岁(四分位间距:1.5,7.5),61%为男性。Pliance X在静止界面压力方面的重测信度接近临床使用可接受水平(ICC = 0.87,n = 27)。Pliance X对于所有未感到痛苦的儿童来说是合适且可接受的。

讨论

需要进一步进行信度测试,包括评分者间信度测试。可以通过使用不同的传感器样式和即时平均压力读数来改善该设备的临床效用,以便能够立即获得关于界面压力的临床有用信息。