Calibration of radiochromic EBT3 film using laser-accelerated protons.

作者信息

Mahmood M Ahsan, Lee Seong Geun, Lee Sang Hwa, Kim Ha-Na, Lee Kitae, Ahmad Izhar, Yang Jeong Moon, Yoon Jin Woo, Lee Hwang Woon, Sung Jae Hee, Lee Seong Ku, Choi Il Woo, Nam Chang Hee

机构信息

Center for Relativistic Laser Science, Institute for Basic Science, Gwangju 61005, Republic of Korea.

Research Center for Ultrafast Science, Korea Atomic Energy Research Institute, Daejeon 34057, Republic of Korea.

出版信息

Rev Sci Instrum. 2021 Feb 1;92(2):023302. doi: 10.1063/5.0031253.

Abstract

We present a proof of principle for onsite calibration of a radiochromic film (EBT3) using CR-39 as an absolute proton-counting detector and laser-accelerated protons as a calibration source. A special detector assembly composed of aluminum range filters, an EBT3 film, and a CR-39 detector is used to expose the EBT3 film with protons in an energy range of 3.65 MeV-5.85 MeV. In our design, the proton beam is divided into small beamlets and their projection images are taken on the EBT3 film and the CR-39 detector by maintaining a certain distance between the two detectors. Owing to the geometrical factor of the configuration and scattering inside the EBT3, the areal number density of protons was kept below the saturation level of the CR-39 detector. We also present a method to relate the number of protons detected on the CR-39 in a narrow energy range to protons with a broad energy spectrum that contribute to the dose deposited in the EBT3 film. The energy spectrum of protons emitted along the target normal direction is simultaneously measured using another CR-39 detector installed in a Thomson parabola spectrometer. The calibration curves for the EBT3 film were obtained in the optical density range of 0.01-0.25 for low dose values of 0.1 Gy-3.0 Gy. Our results are in good agreement with the calibrations of the EBT3 film that are traditionally carried out using conventional accelerators. The method presented here can be further extended for onsite calibration of radiochromic films of other types and for a higher range of dose values.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验